Con
ver
sion Guide
35
HP 8590/HP ESA Spectr
um Analyz
ers Con
version Guide
MKACT
MKACT 1|2|3|4
MKACT?
:CALCulate:MARKer[1]|2|3|4:STATe ON|1
<no SCPI equivalent>
Specifies the active marker.
The HP 8590-Series analyzer outputs data in
the format: 1. The HP ESA analyzer outputs
data in the format: +1.
MKACTV
Makes the current active marker the active
function.
MKBW
Returns the bandwidth at the specified power
level relative to an on-screen marker (if
present) or the signal peak (if no on-screen
marker is present).
MKCF
:CALCulate:MARKer[1]|2|3|4[:SET]:CENTer
Sets the center frequency equal to the marker
frequency and moves the marker to the center
of the screen.
Sets the center frequency equal to the specified
marker frequency, which moves the marker to
the center of the screen.
MKCONT
Resumes the sweep after execution of a
MKSTOP command.
MKD
:CALCulate:MARKer[1]|2|3|4:MODE DELTa
or,
:CALCulate:MARKer[1]|2|3|4:X <param>
Activates the delta marker.
Positions the designated marker on the
assigned trace at the specified X value. The
value is in the x axis units (which is often
frequency or time).
HP 8590-Series
Analyzers
Command(s)
Corresponding SCPI Command(s)
Description/Comments
Summary of Contents for E4401B
Page 6: ...vi ...
Page 14: ...xiv Contents ...
Page 64: ...2 34 Chapter2 HP ESA Status Registers Using the Analyzer Status Registers ...
Page 65: ...3 1 3 Programming Examples ...
Page 151: ...4 1 4 Programming Command Cross References ...
Page 286: ...5 132 Chapter5 Language Reference UNIT Subsystem ...
Page 287: ...6 1 6 HP 8590 HP ESA Spectrum Analyzers Programming Conversion Guide ...
Page 292: ...1 HP 8590 HP ESA Spectrum Analyzers Conversion Guide ...
Page 368: ...7 1 7 Error Messages ...
Page 393: ...7 26 Chapter7 Error Messages 201 to 799 Device Specific Errors ...