![HP E4401B Programmer'S Manual Download Page 328](http://html.mh-extra.com/html/hp/e4401b/e4401b_programmers-manual_161548328.webp)
Con
ver
sion Guide
37
HP 8590/HP ESA Spectr
um Analyz
ers Con
version Guide
MKFCR
MKFCR <freq>
MKFCR AUTO
MKFCR UP|DN
MKFCR?
:CALCulate:MARKer:FCOunt:RESolution <real>
:CALCulate:MARKer:FCOunt:RESolution:AUTO ON|1
:CALCulate:MARKer:FCOunt:RESolution?
Sets the resolution of the marker frequency
counter.
Sets the resolution of the marker frequency
counter. AUTO ON couples the marker counter
resolution to the frequency span.
Sets the resolution of the marker frequency
counter so it is automatically coupled to the
frequency span, generating the fastest
accurate count.
The HP 8590-Series analyzer outputs data in
the format: 1000. The HP ESA analyzer
outputs data in the format: +1000.
MKMIN
:CALCulate:MARKer[1]|2|3|4:MINimum
Moves active marker to minimum signal
detected.
Places the selected marker on the lowest point
on the trace that is assigned to that particular
marker number.
HP 8590-Series
Analyzers
Command(s)
Corresponding SCPI Command(s)
Description/Comments
Summary of Contents for E4401B
Page 6: ...vi ...
Page 14: ...xiv Contents ...
Page 64: ...2 34 Chapter2 HP ESA Status Registers Using the Analyzer Status Registers ...
Page 65: ...3 1 3 Programming Examples ...
Page 151: ...4 1 4 Programming Command Cross References ...
Page 286: ...5 132 Chapter5 Language Reference UNIT Subsystem ...
Page 287: ...6 1 6 HP 8590 HP ESA Spectrum Analyzers Programming Conversion Guide ...
Page 292: ...1 HP 8590 HP ESA Spectrum Analyzers Conversion Guide ...
Page 368: ...7 1 7 Error Messages ...
Page 393: ...7 26 Chapter7 Error Messages 201 to 799 Device Specific Errors ...