5-94
Chapter 5
Language Reference
SENSe Subsystem
SENSe:POWer Subsection
Input Attenuation
[:SENSe]:POWer[:RF]:ATTenuation <rel_ampl>
[:SENSe]:POWer[:RF]:ATTenuation?
Set the input attenuator. This value is set at its auto value if input
attenuation is set to auto.
Factory Preset
and *RST:
10 dB
Range:
HP ESA E4401B, E4411B: 0 to 60 dB
HP ESA E4402B, E4403B: 0 to 75 dB
HP ESA E4404B: 0 to 75 dB
HP ESA E4405B: 0 to 75 dB
HP ESA E4407B, E4408B: 0 to 65dB
Default Unit:
dB
Front Panel
Access:
AMPLITUDE/Y Scale, Attenuation Auto Man
Input Port Attenuator Auto
[:SENSe]:POWer[:RF]:ATTenuation:AUTO OFF|ON|0|1
[:SENSe]:POWer[:RF]:ATTenuation:AUTO?
Select the input port attenuator range to be set either automatically or
manually.
On – Input attenuation is automatically set as determined by the
Reference Level Setting.
Off – Input attenuation is manually set
Factory Preset
and *RST:
On
Range:
HP ESA E4401B, E4411B: 0 to 60 dB
HP ESA E4402B, E4403B: 0 to 75 dB
HP ESA E4404B: 0 to 75 dB
HP ESA E4405B: 0 to 75 dB
HP ESA E4407B, E4408B: 0 to 65dB
Front Panel
Access:
Input, Input Atten
Summary of Contents for E4401B
Page 6: ...vi ...
Page 14: ...xiv Contents ...
Page 64: ...2 34 Chapter2 HP ESA Status Registers Using the Analyzer Status Registers ...
Page 65: ...3 1 3 Programming Examples ...
Page 151: ...4 1 4 Programming Command Cross References ...
Page 286: ...5 132 Chapter5 Language Reference UNIT Subsystem ...
Page 287: ...6 1 6 HP 8590 HP ESA Spectrum Analyzers Programming Conversion Guide ...
Page 292: ...1 HP 8590 HP ESA Spectrum Analyzers Conversion Guide ...
Page 368: ...7 1 7 Error Messages ...
Page 393: ...7 26 Chapter7 Error Messages 201 to 799 Device Specific Errors ...