5-58
Chapter 5
Language Reference
MEASurement Subsystem
Default Unit:
%
Remarks:
The total harmonic distortion precision for SCPI and
the display is three significant digits.
Front Panel
Access:
MEASURE, Harmonic Dist
Return Harmonic Amplitudes
:MEASure:HARMonics:AMPLitude[:ALL]?
Returns a comma-separated list of the amplitudes of the ten harmonics.
Range:
–180 dBm to 70 dBm
Default Unit:
dBm (from fundamental); dBc (all others)
Remarks:
The first value (for the fundamental) is measured in
dBm; the remaining harmonics are measured in dBc. If
fewer than ten harmonics are measured, the value NaN
is returned for any harmonic not measured. The
harmonic amplitude precision for SCPI and the display
is two decimal places.
Front Panel
Access:
MEASURE, Harmonic Dist
Return Harmonic N Amplitude
:MEASure:HARMonics:AMPLitude[n]?
Returns the amplitude of harmonic number n, measured in dBc from
the fundamental, or in dBm if n=1.
Range:
–180 dBm to 70 dBm
Default Unit:
dBc (from fundamental); dBm (n=1)
Remarks:
The harmonic amplitude precision for SCPI and the
display is two decimal places.
Front Panel
Access:
MEASURE, Harmonic Dist
Return Harmonic Frequency List
:MEASure:HARMonics:FREQuency[:ALL]?
Returns a comma-separated list of the frequencies of the ten harmonics,
measured in Hz.
Range:
0 Hz to the maximum frequency range of the analyzer
Default Unit:
Hz
Summary of Contents for E4401B
Page 6: ...vi ...
Page 14: ...xiv Contents ...
Page 64: ...2 34 Chapter2 HP ESA Status Registers Using the Analyzer Status Registers ...
Page 65: ...3 1 3 Programming Examples ...
Page 151: ...4 1 4 Programming Command Cross References ...
Page 286: ...5 132 Chapter5 Language Reference UNIT Subsystem ...
Page 287: ...6 1 6 HP 8590 HP ESA Spectrum Analyzers Programming Conversion Guide ...
Page 292: ...1 HP 8590 HP ESA Spectrum Analyzers Conversion Guide ...
Page 368: ...7 1 7 Error Messages ...
Page 393: ...7 26 Chapter7 Error Messages 201 to 799 Device Specific Errors ...