![HP E4401B Programmer'S Manual Download Page 190](http://html.mh-extra.com/html/hp/e4401b/e4401b_programmers-manual_161548190.webp)
5-36
Chapter 5
Language Reference
CONFigure Subsystem
CONFigure Subsystem
CONFigure subsystem commands apply only to measurements found in
the
MEASURE
menu. These commands stop the current measurement
and set up the instrument for the specified measurement using the
factory default instrument settings.
:CONF:<measurement>
will always set
:INIT:CONT OFF
(single
mode), and also places the measurement in the idle state.
:CONF:SANalyzer
causes the present measurement to exit
(functionally the same as pressing
MEASURE, Meas Off
), and places the
analyzer in base instrument spectrum analyzer state.
The query
:CONFigure?
returns the current measurement name in
quotes.
Configure the Channel Power Measurements
:CONFigure:ACPower
Configure the Channel Power and Density Measurements
:CONFigure:CHPower
Configure the Emission Bandwidth Measurements
:CONFigure:EBWidth
Set Number of Measured Harmonics
:CONFigure:HARMonics[:<number>]
Set the number of harmonics to be measured.
Factory Preset
and *RST:
10
Range:
2 to 10, limited to the maximum frequency range of the
analyzer
Remarks:
This command sets the number of harmonics to
measure before computing the total harmonic
distortion. The minimum number is two (the
fundamental and the second harmonic).
Front Panel
Access:
MEASURE, Harmonic Dist, Meas Setup, Harmonics
Configure the OBW and Transmit Frequency Error
Measurements
:CONFigure:OBWidth
Summary of Contents for E4401B
Page 6: ...vi ...
Page 14: ...xiv Contents ...
Page 64: ...2 34 Chapter2 HP ESA Status Registers Using the Analyzer Status Registers ...
Page 65: ...3 1 3 Programming Examples ...
Page 151: ...4 1 4 Programming Command Cross References ...
Page 286: ...5 132 Chapter5 Language Reference UNIT Subsystem ...
Page 287: ...6 1 6 HP 8590 HP ESA Spectrum Analyzers Programming Conversion Guide ...
Page 292: ...1 HP 8590 HP ESA Spectrum Analyzers Conversion Guide ...
Page 368: ...7 1 7 Error Messages ...
Page 393: ...7 26 Chapter7 Error Messages 201 to 799 Device Specific Errors ...