ver
sion
Guide
HP 8590/HP ESA Spectr
um Analyz
ers Con
version Guide
AMB
Subtracts trace B from trace A and sends the
result to trace A during every sweep of the
spectrum analyzer.
AMBPL
Subtracts trace B from trace A, adds the
display line value to the difference, and sends
the result to trace A during every sweep of the
spectrum analyzer.
AMPCOR
AMPCOR (data)
AMPCOR
OFF/ON
AMPCOR?
[SENSE:]CORRection:CSET[1]|2|3|4:DATA <freq>,
<rel_ampl>{,<freq>, <rel_ampl>}
[:SENSe]:CORRection:CSET:ALL[:STATe] OFF|ON|0|1
[:SENSe]:CORRection:CSET[1]|2|3|4[:STATe]
OFF|ON|0|1
[:SENSe]:CORRection:CSET[1]|2|3|4:DATA?
Applies amplitude correction at specified
frequencies.
Units and spaces are not allowed in SCPI. The
separator must be a comma (,) and the
terminator must be a semicolon (;). Frequency
and amplitude values must be entered in Hz
and dB.
The data format for the command and query is
always TDF P.
The HP 8590-Series analyzer returns data in
the format: -57.71, -58.12, -56.87. The HP ESA
analyzer returns data in the format:
-5.46001, -5.44001,
-5.47001. This is an example of IEEE
NR3 numeric response data.
HP 8590-Series
Analyzers
Command(s)
Corresponding SCPI Command(s)
Description/Comments
Summary of Contents for E4401B
Page 6: ...vi ...
Page 14: ...xiv Contents ...
Page 64: ...2 34 Chapter2 HP ESA Status Registers Using the Analyzer Status Registers ...
Page 65: ...3 1 3 Programming Examples ...
Page 151: ...4 1 4 Programming Command Cross References ...
Page 286: ...5 132 Chapter5 Language Reference UNIT Subsystem ...
Page 287: ...6 1 6 HP 8590 HP ESA Spectrum Analyzers Programming Conversion Guide ...
Page 292: ...1 HP 8590 HP ESA Spectrum Analyzers Conversion Guide ...
Page 368: ...7 1 7 Error Messages ...
Page 393: ...7 26 Chapter7 Error Messages 201 to 799 Device Specific Errors ...