5-10
Chapter 5
Language Reference
CALCulate Subsystem
CALCulate Subsystem
This subsystem is used to perform post-acquisition data processing. In
effect, the collection of new data triggers the CALCulate subsystem. In
this instrument, the primary functions in this subsystem are markers
and limits.
NdBpoints
:CALCulate:BWIDth|BANDwidth:NDB <rel_ampl>
:CALCulate:BWIDth|BANDwidth:NDB?
Selects the power level, below the peak of the signal, at which the
signal bandwidth will be measured by the markers.
:CALCulate:BWIDth|BANDwidth[:STATe]
must be ON.
NOTE
To query the result of NdBpoints, use the command
:CALCulate:BWIDth|BANDwidth:RESult?
Factory Preset
and *RST:
–3 dB
Range:
–80 dB to –1 dB
Default Unit:
dB
Remarks:
Refer to
:CALCulate:BWIDth|BANDwidth[:STATe]
for
an explanation of this marker function.
Front Panel
Access:
Search, N dB Points
NdBresults
:CALCulate:BWIDth|BANDwidth:RESult?
Returns the measured bandwidth at the power level defined by
:CALCulate:BWIDth:NDB
. –100 is returned if
:CALCulate:BWIDth|BANDwidth[:STATe]
is off, or when a result is
not available.
Range:
Real value less than the current frequency span
Default Unit:
Hz
Remarks:
Refer to
CALCulate:BWIDth|BANDwidth[:STATe]
for an
explanation of this marker function.
Front Panel
Access:
Search, N dB Points
Summary of Contents for E4401B
Page 6: ...vi ...
Page 14: ...xiv Contents ...
Page 64: ...2 34 Chapter2 HP ESA Status Registers Using the Analyzer Status Registers ...
Page 65: ...3 1 3 Programming Examples ...
Page 151: ...4 1 4 Programming Command Cross References ...
Page 286: ...5 132 Chapter5 Language Reference UNIT Subsystem ...
Page 287: ...6 1 6 HP 8590 HP ESA Spectrum Analyzers Programming Conversion Guide ...
Page 292: ...1 HP 8590 HP ESA Spectrum Analyzers Conversion Guide ...
Page 368: ...7 1 7 Error Messages ...
Page 393: ...7 26 Chapter7 Error Messages 201 to 799 Device Specific Errors ...