3-50
Chapter 3
Programming Examples
Entering Amplitude Correction Data
Entering Amplitude Correction Data
This C programming example does the following:
• Open an HP-IB session at address 18.
• Clear the analyzer.
• Reset the analyzer.
• Set the analyzer stop frequency to 1.5 GHz.
• Set the input port to the 50 MHz amplitude reference.
• Enter amplitude correction frequency/amplitude
pairs:100 MHz/5 dB, 1 GHz/
−
5 dB, 1.5MHz/10 dB.
• Activate amplitude correction.
• Store the array (frequency, amplitude) value into a two-dimensional
real array.
• Query the analyzer for the amplitude correction factors.
• Remove the "," token from the amplitude correction for analyzing
data.
• Display the contents of the array.
• Close the session.
Example:
#include <stdio.h>
#include <stdlib.h>
#include <math.h>
#include <conio.h>
#include <ctype.h>
#include <string.h>
#include "visa.h"
#define hpESA_IDN_E4401B
"Hewlett-Packard, E4401B"
#define hpESA_IDN_E4411B
"Hewlett-Packard, E4411B"
Visession defaultRM, viESA;
ViStatus
errStatus;
ViChar
cIdBuff[256]={0};
char
cEnter = 0;
int
iResult = 0;
/*Set the input port to the 50 MHz amplitude reference*/
void Route50MHzSignal()
Summary of Contents for E4401B
Page 6: ...vi ...
Page 14: ...xiv Contents ...
Page 64: ...2 34 Chapter2 HP ESA Status Registers Using the Analyzer Status Registers ...
Page 65: ...3 1 3 Programming Examples ...
Page 151: ...4 1 4 Programming Command Cross References ...
Page 286: ...5 132 Chapter5 Language Reference UNIT Subsystem ...
Page 287: ...6 1 6 HP 8590 HP ESA Spectrum Analyzers Programming Conversion Guide ...
Page 292: ...1 HP 8590 HP ESA Spectrum Analyzers Conversion Guide ...
Page 368: ...7 1 7 Error Messages ...
Page 393: ...7 26 Chapter7 Error Messages 201 to 799 Device Specific Errors ...