![HP E4401B Programmer'S Manual Download Page 233](http://html.mh-extra.com/html/hp/e4401b/e4401b_programmers-manual_161548233.webp)
Chapter 5
5-79
Language Reference
SENSe Subsystem
specified correction set.
Remarks:
Logarithmic frequency scale corrections are linearly
interpolated between correction points with respect to
the logarithm of the frequency. Linear frequency scale
corrections are interpolated along straight lines,
connecting adjacent points on a linear scale.
Front Panel
Access:
AMPLITUDE/Y Scale, Corrections, Modify, Freq Interp Log
Lin
Perform Amplitude Correction
[:SENSe]:CORRection:CSET[1]|2|3|4[:STATe] OFF|ON|0|1
[:SENSe]:CORRection:CSET[1]|2|3|4[:STATe]?
Turns the amplitude correction function on or off for the given set.
NOTE
[:SENSe]:CORRection:CSET:ALL[:STATe]
must be on for this command
to function.
Factory Preset
and *RST:
Off
Remarks:
CSET number equivalents to front panel access
definitions are as follows:
CSET1 is Antenna
CSET2 is Cable
CSET3 is Other
CSET4 is User
Front Panel
Access:
AMPLITUDE/Y Scale, Corrections, Modify, Correction On
Off
Input Impedance Correction
[:SENSe]:CORRection:IMPedance[:INPut][:MAGNitude] <number>
[:SENSe]:CORRection:IMPedance[:INPut][:MAGNitude]?
Amplitude correction is applied to the display data to adjust for
measurement situations where the unit under test has a different
impedance than the 50
Ω
input impedance of the analyzer. Some HP
ESA analyzers have Option 1DP, 75
Ω
input. In this case, you may want
to convert the data to make measurements in a 50
Ω
system.
Factory Preset
and *RST:
The factory default is the input impedance of the
analyzer.
Summary of Contents for E4401B
Page 6: ...vi ...
Page 14: ...xiv Contents ...
Page 64: ...2 34 Chapter2 HP ESA Status Registers Using the Analyzer Status Registers ...
Page 65: ...3 1 3 Programming Examples ...
Page 151: ...4 1 4 Programming Command Cross References ...
Page 286: ...5 132 Chapter5 Language Reference UNIT Subsystem ...
Page 287: ...6 1 6 HP 8590 HP ESA Spectrum Analyzers Programming Conversion Guide ...
Page 292: ...1 HP 8590 HP ESA Spectrum Analyzers Conversion Guide ...
Page 368: ...7 1 7 Error Messages ...
Page 393: ...7 26 Chapter7 Error Messages 201 to 799 Device Specific Errors ...