5 Maintenance
57
5.3
Replacement of parts
If a filter or packing event occurred, replace the part.
5.3.1 How to replace the filters and packing
Note
Filters are used to protect the analyzer. When used for a long period of time, the filter elements will
become clogged resulting in a decrease in the flow rate of the sample gas.
1.
Remove the screws of the gas connectors on both sides.
2.
Replace the filters and packing.
Fig. 67 Rear panel filter section
3.
Refer to "Resetting the passed times" (page 42) and then reset the limit time.
Packing
Filter