6 Troubleshooting
63
6.3.2 Types of alarms and remedial actions
Table 29 Types of alarms and remedial actions
Name
Description
Check details
Remedial action
Contact output
SETTING
LOAD
The settings could
not be loaded at
startup.
―
Contact HORIBA or local
representative.
OFF during normal
operation (closed
during operation)
(page 66)
CAN
COMM.
(ID1)
An error is
occurring with CAN
communication
between the circuit
boards inside the
analyzer.
―
Contact HORIBA or local
representative.
OFF during normal
operation (closed
during operation)
(page 66)
CAN
COMM.
(ID2)
An error is
occurring with CAN
communication
between the circuit
boards inside the
analyzer.
―
Contact HORIBA or local
representative.
OFF during normal
operation (closed
during operation)
(page 66)
ADC
The AD conversion
value of the
detector signal is
out of the specified
range.
Voltage value of
each component
(ANALOG INPUT
screen)
Check whether the sample
gas concentration is too
high.
OFF during normal
operation (closed
during operation)
(page 66)
CELL
TEMP
The cell
temperature
exceeds the
specified value.
Cell temperature
(ANALOG INPUT
screen)
Check whether the ambient
temperature meets the
specifications.
OFF during normal
operation (closed
during operation)
(page 66)
SAMPLE
PRS.
The pressure of the
sample gas is
decreasing.
Sample pressure
switch
(ANALOG INPUT
screen)
Check the pressure and
flow rate of the sample gas.
Flow rate error (closed
during operation) ON
OFF during normal
operation (closed
during operation)
(page 66)
REFEREN
CE PRS.
The pressure of the
reference gas is
decreasing.
Reference
pressure switch
(ANALOG INPUT
screen)
Check the pressure and
flow rate of the reference
gas.
Flow rate error (closed
during operation)
Signal OFF during
normal operation
(page 66)
ADC PCB
TEMP
The ADC board
temperature
exceeds the
specified value.
ADC board
temperature
(ANALOG INPUT
screen)
Check whether the ambient
temperature meets the
specifications.
OFF during normal
operation (closed
during operation)
(page 66)
ATM.
PRESSUR
E
The atmospheric
pressure exceeds
the specified value.
Atmospheric
pressure
(ANALOG INPUT
screen)
Contact HORIBA or local
representative.
OFF during normal
operation (closed
during operation)
(page 66)
SV
MISSING
The solenoid valve
of the analysis part
is not energized.
―
Contact HORIBA or local
representative.
OFF during normal
operation (closed
during operation)
(page 66)