8 Reference
74
How to fix the fixtures
1.
Change the suspension position for the springs.
Change the suspension position for the springs from the top level to the bottom level.
2.
Fix the solenoid valve unit fixture.
Remove the screw attached to the solenoid valve unit. Loosen the screw at the bottom of
the fixture, rotate the fixture half a rotation (turn it to the opposite side) and secure with a
screw. Attach the screw you removed from the solenoid valve unit to the top of the fixture
to fix it to the solenoid valve unit.
3.
Fix the analysis base to the chassis.
Remove the screws attached to part A of the analysis part base and attach them to the 4
places on the analysis part base to fix it to the chassis.
Part A
Analysis part base