2 Preparations
7
2
Preparations
This chapter specifies the installation conditions and describes the tube and cable
connections.
2.1
Contact for installation
2.2
Installation conditions
This analyzer is designed to be used under ordinary environmental conditions so
consideration has not been given to use under special conditions.
Install the analyzer in a location that meets the following conditions.
Location sheltered from rain and water drops
Location with an ambient temperature of 0°C to 40°C that is not subject to sudden
temperature changes of 5
C or more.
Location not directly exposed to heat from direct sunlight or a heater
Location with little dust
Location with little vibration
Location free from powerful electric and magnetic fields
Location where corrosive gas is not present in the atmosphere
Location with relative humidity of 85% or less
Location where maintenance space can be provided
Maintenance needs to be performed from the front, rear and top. Furthermore, a
ventilation fan is provided to prevent the temperature inside the case from rising. Provide
space from the walls for the ventilation fan.
Location indoors
Supply voltage fluctuations: ±10% of reference voltage (up to 250 V)
Power frequency: within ±1% of reference frequency
Installation environment of main power source conforming to overvoltage category II and
pollution degree 2
Altitude of 3000 m or less
Placed on a level surface or fixed to a rack
Air inlets and exhaust outlet are not blocked
Leave a space of at least 10 mm from the wall.
Manufacturer: HORIBA,
Ltd.
2 Miyanohigashi, Kisshoin Minami-ku, Kyoto 601-8510 Japan
Summary of Contents for GA-370
Page 1: ...Trace Gas Monitor GA 370 Instruction Manual eng CODE GZ0000456936B...
Page 92: ......
Page 93: ......
Page 94: ......
Page 95: ...2 Miyanohigashi Kisshoin Minami ku Kyoto 601 8510 Japan http www horiba com...