![horiba GA-370 Instruction Manual Download Page 62](http://html1.mh-extra.com/html/horiba/ga-370/ga-370_instruction-manual_2162216062.webp)
4 Various Functions
46
Note
If more than 8 items exist, the
and
buttons are displayed so that you can change the
page screen.
Up to 60 history items can be recorded and displayed for each component. When 60 items will
be exceeded, the oldest item is deleted and the new item is recorded.
4.4.5 Event history check
Check the latest event history (up to 100 items).
Fig. 51 EVENT HISTORY screen
The displayed information is as follows.
Table 22 Event History display items
Note
When the power is turned on, the [ANALYZER START] information is automatically saved to the
event history (the EVENTS screen of the MEASUREMENT screen is not displayed). It is
information for checking the time the analyzer started, and has no effect on performance.
DATA MENU
[EVENT HISTORY]
Item
Description
[TIME AND DATE]
Time and date the event occurred
[EVENT]
Event that occurred
[COND]
Event status
ON: Event occurring/started
OFF: Event canceled/ended