3 Operations
21
Span gas concentration settings
Set the span gas concentration to use for calibration.
Operating procedure
1.
Press the [SPAN GAS CONC.] button in the CALIBRATION MENU screen.
The SPAN GAS CONC. screen appears.
2.
Press the span gas concentration button of the component to be calibrated.
Fig. 16 SPAN GAS CONC. screen
3.
Enter the span gas concentration value and then press the [Enter] button.
The setting is updated and the SPAN GAS CONC. screen reappears.
Fig. 17 Span gas concentration value input screen