Vertex M
TM
24-Point Continuous Monitor
Section 2 - Installation
2-6
2.6.1 Connecting AC Power
AC Source Requirements:
• Operating Voltage: 110 or 230 VAC ± 10%
(under load) @ 50/60 Hz; 15 Amps maximum,
single phase.
The Vertex M System requires a dedicated AC circuit
rated at either 110 or 230 volts, 50/60 Hz, 15 Amp single
phase. Line voltage should fluctuate no more than
± 10%. The external switch must be clearly labeled and
installed in accordance with local electrical codes. Input
power cable should be #14 AWG minimum. The safety
ground wire must be the same or larger gauge as the
line wires. Connect AC power connection to the three-
position terminal block in the side panel of the rack.
See Figure 2-2.
Neutral
Ground
Line
AC Input
Figure 2-2. AC Power Connection
2.6.2 Power On/Off
An internal rack power switch is located behind the
door.
After performing self-diagnostics, the Vertex M System
main screen opens and the system returns to the same
state it was in prior to power down.
Rack Power Switch
Analyzer / Pyrolyzer
Power Switches and Indicators
Figure 2-3. Rack Power Switch
WARNING
Hazardous voltages may exist at the Alarm
Contacts in this unit with the power switch
turned off. Insure power is disconnected at
the source prior to servicing alarm contacts.
2.7 Data Acquisition System
The data acquisition computer or DAq is the main
computer in the Vertex M System. System display and
operator control are through an LCD touch screen with
on-screen keyboard or the external keyboard on top
of the unit.
Summary of Contents for Vertex M
Page 8: ...1 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 1 Introduction ...
Page 25: ...Vertex MTM 24 Point Continuous Monitor Section 1 Introduction 1 18 ...
Page 26: ...2 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 2 Installation ...
Page 35: ...Vertex MTM 24 Point Continuous Monitor Section 2 Installation 2 10 ...
Page 36: ...3 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 3 Startup ...
Page 40: ...Vertex MTM 24 Point Continuous Monitor Section 3 Startup 3 5 Figure 3 1 Vertex M Main Screen ...
Page 77: ...4 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 4 Operation ...
Page 93: ...Vertex MTM 24 Point Continuous Monitor Section 4 Operation 4 17 Figure 4 9 Filter Options ...
Page 109: ...Vertex MTM 24 Point Continuous Monitor Section 4 Operation 4 33 ...
Page 125: ...5 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 5 Maintenance ...
Page 134: ...Vertex MTM 24 Point Continuous Monitor Section 5 Maintenance 5 10 ...
Page 140: ...Vertex MTM 24 Point Continuous Monitor Section 5 Maintenance 5 16 ...
Page 141: ...6 1 Vertex M TM Technical Handbook 6 Troubleshooting ...
Page 164: ...Section 6 Troubleshooting 6 24 ...
Page 166: ...Vertex MTM 24 Point Continuous Monitor Appendix A Installation Drawings A 2 A 1 Introduction ...
Page 170: ...Vertex MTM 24 Point Continuous Monitor Appendix A Installation Drawings A 6 ...
Page 173: ...Vertex MTM 24 Point Continuous Monitor Appendix A Installation Drawings A 9 ...
Page 174: ...B 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook A Specifications ...
Page 178: ...C 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook A Detectable Gases ...
Page 193: ...Vertex MTM 24 Point Continuous Monitor Appendix A Replacement and Consumable Items D 4 ...
Page 240: ...Vertex MTM 24 Point Continuous Monitor Appendix A 4 20mA Analog Output Option G 4 ...
Page 250: ...Vertex MTM 24 Point Continuous Monitor Appendix A Warranty Statement I 4 ...