Vertex M
TM
24-Point Continuous Monitor
Section 4 - Operation
4-6
Pump Status Indicator
The Vertex Display includes pump status indicators in
the top right corner. The bottom row indicates which
of the pumps is currently operating. The indicator will
display “ON” with a green background if the pump is
operating and “OFF” on a white background if the pump
is not running.
The top row indicates what is known about pump
health. The indicator will display “GOOD” with a green
background if the pump has successfully provided
proper levels of system vacuum. If the Indicator
displays “BAD” on a red background, this indicates that
pump related faults 112 or 219 have been generated.
The indicator may display “UNKNOWN” on a white
background after software is installed or after analyzers
are added or moved into a different location.
Pumps in the “UNKNOWN” or “BAD” state change to
the “GOOD” state only when successfully used during
gas monitoring. After a failed pump has been repaired
or replaced, it can be exercised by pressing “PUMP
ALTERNATE” button while monitoring. See
Section
4.6.2 Flow Calibration
. If successful, the pump state
will change to “GOOD”.
Honeywell Analytics recommends that pumps be
alternated periodically to insure availability according
to your facility’s schedule.
Pump Health
Indication
Pump Operation
Indication
Pump Status Indicator
Summary of Contents for Vertex M
Page 8: ...1 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 1 Introduction ...
Page 25: ...Vertex MTM 24 Point Continuous Monitor Section 1 Introduction 1 18 ...
Page 26: ...2 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 2 Installation ...
Page 35: ...Vertex MTM 24 Point Continuous Monitor Section 2 Installation 2 10 ...
Page 36: ...3 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 3 Startup ...
Page 40: ...Vertex MTM 24 Point Continuous Monitor Section 3 Startup 3 5 Figure 3 1 Vertex M Main Screen ...
Page 77: ...4 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 4 Operation ...
Page 93: ...Vertex MTM 24 Point Continuous Monitor Section 4 Operation 4 17 Figure 4 9 Filter Options ...
Page 109: ...Vertex MTM 24 Point Continuous Monitor Section 4 Operation 4 33 ...
Page 125: ...5 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 5 Maintenance ...
Page 134: ...Vertex MTM 24 Point Continuous Monitor Section 5 Maintenance 5 10 ...
Page 140: ...Vertex MTM 24 Point Continuous Monitor Section 5 Maintenance 5 16 ...
Page 141: ...6 1 Vertex M TM Technical Handbook 6 Troubleshooting ...
Page 164: ...Section 6 Troubleshooting 6 24 ...
Page 166: ...Vertex MTM 24 Point Continuous Monitor Appendix A Installation Drawings A 2 A 1 Introduction ...
Page 170: ...Vertex MTM 24 Point Continuous Monitor Appendix A Installation Drawings A 6 ...
Page 173: ...Vertex MTM 24 Point Continuous Monitor Appendix A Installation Drawings A 9 ...
Page 174: ...B 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook A Specifications ...
Page 178: ...C 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook A Detectable Gases ...
Page 193: ...Vertex MTM 24 Point Continuous Monitor Appendix A Replacement and Consumable Items D 4 ...
Page 240: ...Vertex MTM 24 Point Continuous Monitor Appendix A 4 20mA Analog Output Option G 4 ...
Page 250: ...Vertex MTM 24 Point Continuous Monitor Appendix A Warranty Statement I 4 ...