Event
Number
Event List
Desc. (Short)
Event History
Description
Data Field
Problem
Possible Cause
Recovery
241
OPT REF ADJ
TIMEOUT
Optics reference
adjustment
timeout
Error code
Optics block self-
adjustment timeout.
Cycle power to analyzer.
Contact Honeywell service.
270
COM AZ # #
BROKEN
OPC To Az ##
Communica-tions
Broken
Analyzer not
responding to OPC
driver polling request
Analyzer powered down
Check switch on power
module
Ethernet cable not
connected
Check Ethernet
connection at the back of
the analyzer
Move cable to different
port on Ethernet hub
301
PLC-COM
BROKEN
PLC-Com
Broken
PLC fails to respond to
DAq communication
Cable connection
problem
Check serial
communications on DAq
COM2
Check serial cable on
PLC 1747-KE
See
Section 1.2.7 Data
Acquisition Computer
(rear)
See
Section 1.2.9 Main
PLC
Loss of PLC power
Check power connections
for PLC
302
AZ-BAD PROF
ID
AZ has wrong
profile loaded
Profile ID in analyzer
does not match current
configuration in use
Configuration not properly
loaded
Reinstall current
Configuration Profile
303
PLC-BAD
PROF ID
PLC Has Wrong
Profile loaded
Profile ID in PLC does
not match current
configuration in use
Configuration not properly
loaded
Reinstall current
Configuration Profile
304
PLC SYSTEM
FAULT
PLC System
Fault
PLC Error
Code
A fault has occurred on
the PLC that prevents
the PLC program from
running
Faulty or missing PLC
module
PLC is not properly
configured
Contact Honeywell
Analytics Service
305
DATA LOG
FAILED
Unable To Log
Conc Data
Failure to store
concentration data in
the database
Local database file has
reached size limit (1GB
maximum size for Access
database).
Enable automatic
database purge or adjust
database purge settings
to purge old data
413
CMD ERR AZ
#-#
Cmd Error Az #-#
Analyzer not
responding DAq
Communica-tions
Cable connections
Check Ethernet
connections
Analyzer missing or
powered down
Remove analyzer from
Configuration Profile
Summary of Contents for Vertex M
Page 8: ...1 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 1 Introduction ...
Page 25: ...Vertex MTM 24 Point Continuous Monitor Section 1 Introduction 1 18 ...
Page 26: ...2 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 2 Installation ...
Page 35: ...Vertex MTM 24 Point Continuous Monitor Section 2 Installation 2 10 ...
Page 36: ...3 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 3 Startup ...
Page 40: ...Vertex MTM 24 Point Continuous Monitor Section 3 Startup 3 5 Figure 3 1 Vertex M Main Screen ...
Page 77: ...4 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 4 Operation ...
Page 93: ...Vertex MTM 24 Point Continuous Monitor Section 4 Operation 4 17 Figure 4 9 Filter Options ...
Page 109: ...Vertex MTM 24 Point Continuous Monitor Section 4 Operation 4 33 ...
Page 125: ...5 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 5 Maintenance ...
Page 134: ...Vertex MTM 24 Point Continuous Monitor Section 5 Maintenance 5 10 ...
Page 140: ...Vertex MTM 24 Point Continuous Monitor Section 5 Maintenance 5 16 ...
Page 141: ...6 1 Vertex M TM Technical Handbook 6 Troubleshooting ...
Page 164: ...Section 6 Troubleshooting 6 24 ...
Page 166: ...Vertex MTM 24 Point Continuous Monitor Appendix A Installation Drawings A 2 A 1 Introduction ...
Page 170: ...Vertex MTM 24 Point Continuous Monitor Appendix A Installation Drawings A 6 ...
Page 173: ...Vertex MTM 24 Point Continuous Monitor Appendix A Installation Drawings A 9 ...
Page 174: ...B 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook A Specifications ...
Page 178: ...C 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook A Detectable Gases ...
Page 193: ...Vertex MTM 24 Point Continuous Monitor Appendix A Replacement and Consumable Items D 4 ...
Page 240: ...Vertex MTM 24 Point Continuous Monitor Appendix A 4 20mA Analog Output Option G 4 ...
Page 250: ...Vertex MTM 24 Point Continuous Monitor Appendix A Warranty Statement I 4 ...