Vertex M
TM
24-Point Continuous Monitor
Appendix A - Detectable Gases
C-6
Gas Name1
TLV
LAL
LDL
Default
Alarm
Level 1
Default
Alarm
Level 2
Range
Alarm Setting
Initial
Analysis
Period
(second)
Time to 1 TLV
(Alarm @ 2 TLV
Concentration
10-ft. Sample
Line)
Chemcassette
Part
Number
Hydrogen Chloride
(HCl)
2 ppm
0.5 ppm
0.5 ppm
2 ppm
4 ppm
0-15 ppm
0.5-9.9 ppm
10-19.9 ppm
20-50 ppm
30
20
15
<25 sec
1295-0225
Hydrogen Chloride
(HCl)
Low Level
2 ppm
100 ppb
80 ppb
250 ppb
500 ppb
0-8000 ppb
100-499 ppb
500-999 ppb
1000-2999 ppb
3000-8000 ppb
300
120
60
30
<30 sec
1295-0225
Hydrogen Chloride XP
(HCl)
2 ppm
0.2 ppm
0.2 ppm
2 ppm
4 ppm
0-20 ppm
0.2-0.9 ppm
1-3.9 ppm
4-20 ppm
60
30
20
<25 sec
1295-0507
Hydrogen Chloride XP4
(HCl)
2 ppm
0.2 ppm
0.2 ppm
2 ppm
4 ppm
0-20 ppm
0.2-0.9 pm
1-3.9 ppm
4-20 ppm
60
30
20
<25 sec
1257-9310
Hydrogen Chloride XP4
(HCl)
Low Level
2 ppm
30 ppb
20 ppb
200 ppb
400 ppb
0-2000 ppb
30-199 ppb
200-399 ppb
400-2000 ppb
240
150
90
<95 sec
(Alarm @ 200
ppb with 400
ppb HCl gas)
1257-9310
Hydrogen Chloride XP
(HCl)
Low Level
2 ppm
30 ppb
20 ppb
200 ppb
400 ppb
0-2000 ppb
30-199 ppb
200-399 ppb
400-2000 ppb
240
150
90
<95 sec
(Alarm @ 200
ppb with 400
ppb HCl gas)
1295-0507
Hydrogen Cyanide
(HCN)
4.7 ppm
1 ppm
0.5 ppm
4.7 ppm
9.4 ppm
0-30 ppm
1-9.9 ppm
10-19.9 ppm
20-30 ppm
30
20
15
<25 sec
1295-0222
Hydrogen Fluoride
(HF)
2 ppm
0.4 ppm
0.4 ppm
2 ppm
4 ppm
0-30 ppm
0.4-2.9 ppm
3-5.9 ppm
6-11.9 ppm
12-30 ppm
120
60
30
15
<45 sec
1295-0225
Hydrogen Fluoride XP (HF)
2 ppm
0.4 ppm
0.4 ppm
2 ppm
4 ppm
0-20 ppm
0.4-0.9 ppm
1-3.9 ppm
4-20 ppm
240
90
60
<50 sec
1295-0507
TLV - Threshold Limit Value LAL - Lowest Alarm Level LDL - Lower Detectable Limit
1. “-1” calibrations allow more stain development but will extend response times
2. Minimum Sample Time
Summary of Contents for Vertex M
Page 8: ...1 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 1 Introduction ...
Page 25: ...Vertex MTM 24 Point Continuous Monitor Section 1 Introduction 1 18 ...
Page 26: ...2 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 2 Installation ...
Page 35: ...Vertex MTM 24 Point Continuous Monitor Section 2 Installation 2 10 ...
Page 36: ...3 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 3 Startup ...
Page 40: ...Vertex MTM 24 Point Continuous Monitor Section 3 Startup 3 5 Figure 3 1 Vertex M Main Screen ...
Page 77: ...4 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 4 Operation ...
Page 93: ...Vertex MTM 24 Point Continuous Monitor Section 4 Operation 4 17 Figure 4 9 Filter Options ...
Page 109: ...Vertex MTM 24 Point Continuous Monitor Section 4 Operation 4 33 ...
Page 125: ...5 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 5 Maintenance ...
Page 134: ...Vertex MTM 24 Point Continuous Monitor Section 5 Maintenance 5 10 ...
Page 140: ...Vertex MTM 24 Point Continuous Monitor Section 5 Maintenance 5 16 ...
Page 141: ...6 1 Vertex M TM Technical Handbook 6 Troubleshooting ...
Page 164: ...Section 6 Troubleshooting 6 24 ...
Page 166: ...Vertex MTM 24 Point Continuous Monitor Appendix A Installation Drawings A 2 A 1 Introduction ...
Page 170: ...Vertex MTM 24 Point Continuous Monitor Appendix A Installation Drawings A 6 ...
Page 173: ...Vertex MTM 24 Point Continuous Monitor Appendix A Installation Drawings A 9 ...
Page 174: ...B 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook A Specifications ...
Page 178: ...C 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook A Detectable Gases ...
Page 193: ...Vertex MTM 24 Point Continuous Monitor Appendix A Replacement and Consumable Items D 4 ...
Page 240: ...Vertex MTM 24 Point Continuous Monitor Appendix A 4 20mA Analog Output Option G 4 ...
Page 250: ...Vertex MTM 24 Point Continuous Monitor Appendix A Warranty Statement I 4 ...