Section 6 - Troubleshooting
6-14
Event
Number
Event List
Desc. (Short)
Event History
Description
Data Field
Problem
Possible Cause
Recovery
228
NO PYRO
MAINS PWR
No Pyrolyzer
Mains Power
1
No mains power at
points 5-8
Internal thermal fuse
opened
Check cooling fan,
replace thermal fuse
2
No mains power at
points 1-4
3
No mains power at all 8
points
Pyro mains switch not on
Circuit breaker tripped
Power connector
disconnected
Defective transformer
Turn on pyro mains switch
Breaker is self resetting
Check connection on
back of Analyzer
Replace transformer
229
OPTICS CFG ERR
Invalid Optics
Configuration
Error code
Microprocessor in optics
block reports problem.
Chemleader read failed
Reread Chemleader.
Cycle power to analyzer.
Update software in optics
block.
230
COM FAIL TO
PLC
No
Communication
From PLC To Az
0.0
No communication from
PLC
Communication has
timed out to PLC
Cycle power to the
analyzers and reload
Configuration Profile
Check FLT light on PLC
Connection at rear of
analyzer
Check 25-pin,
multifunction cable at the
rear of the analyzer
Check cable connections
to the PLC
Check cable connection
at chassis from carrier
231
DRV ADJ FAIL
AUTO LED
Adjustment
Failed
Optics counts LED adjustment failed
Tape incorrectly loaded
Reload and retry tape
load
Dirty optics block
Service or replace
analyzer
Bad optics PCB set
Service or replace
analyzer
Bad RFID tag
Load new Chemcassette
232
BAD
Q-FACTOR
Q-Factor Out Of
Range
Q-Factor
Q-Factor out of range
Chemcassette leader
not tight or improperly
positioned during white to
light gray calibration
Reload Chemcassette
Bad RFID tag
Load new Chemcassette
Dirty optics block
Clean and recalibrate
Bad optics PCB set
Service or replace
analyzer
Summary of Contents for Vertex M
Page 8: ...1 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 1 Introduction ...
Page 25: ...Vertex MTM 24 Point Continuous Monitor Section 1 Introduction 1 18 ...
Page 26: ...2 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 2 Installation ...
Page 35: ...Vertex MTM 24 Point Continuous Monitor Section 2 Installation 2 10 ...
Page 36: ...3 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 3 Startup ...
Page 40: ...Vertex MTM 24 Point Continuous Monitor Section 3 Startup 3 5 Figure 3 1 Vertex M Main Screen ...
Page 77: ...4 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 4 Operation ...
Page 93: ...Vertex MTM 24 Point Continuous Monitor Section 4 Operation 4 17 Figure 4 9 Filter Options ...
Page 109: ...Vertex MTM 24 Point Continuous Monitor Section 4 Operation 4 33 ...
Page 125: ...5 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 5 Maintenance ...
Page 134: ...Vertex MTM 24 Point Continuous Monitor Section 5 Maintenance 5 10 ...
Page 140: ...Vertex MTM 24 Point Continuous Monitor Section 5 Maintenance 5 16 ...
Page 141: ...6 1 Vertex M TM Technical Handbook 6 Troubleshooting ...
Page 164: ...Section 6 Troubleshooting 6 24 ...
Page 166: ...Vertex MTM 24 Point Continuous Monitor Appendix A Installation Drawings A 2 A 1 Introduction ...
Page 170: ...Vertex MTM 24 Point Continuous Monitor Appendix A Installation Drawings A 6 ...
Page 173: ...Vertex MTM 24 Point Continuous Monitor Appendix A Installation Drawings A 9 ...
Page 174: ...B 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook A Specifications ...
Page 178: ...C 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook A Detectable Gases ...
Page 193: ...Vertex MTM 24 Point Continuous Monitor Appendix A Replacement and Consumable Items D 4 ...
Page 240: ...Vertex MTM 24 Point Continuous Monitor Appendix A 4 20mA Analog Output Option G 4 ...
Page 250: ...Vertex MTM 24 Point Continuous Monitor Appendix A Warranty Statement I 4 ...