6.3 Maintenance Faults
A maintenance fault indicates the Vertex M System requires attention but is continuing to monitor. When a
maintenance fault occurs, the following actions take place within the Vertex M System:
• The LED indicator flashes yellow on the analyzer with the fault. See
Table 6-1. Analyzer Status LEDs
.
• A yellow fault indicator appears on the main screen
• The event list and event history are updated
• Maintenance Fault relays associated with this analyzer will activate
Event
Number
Event List Desc.
(Short)
Event History
Description
Data Field
Problem
Possible Cause
Recovery
101
FLOW ADJ ERR
LOW
Low Flow After
Adjustment
cc/min. of
error
Point Flow reading can
not reach 150 cc/min.
Verify range of valve by
performing Auto Balance
calibration
Identify possible cause if
maximum flow is below
240 cc/min.
Condensation
Check internal lines for
moisture
Purge as required
Internal proportional
valve filter clogged
Replace particulate filter
at manifold
Proportional valve failure Replace Valve
Supply vacuum
insufficient (less than 7
in. Hg)
Plug pneumatic
connector in unused slots
Exhaust tubing restricted
Service or switch pumps
Excessive point vacuum
Sample line restricted
End of line filter plugged
Too many analyzers
sharing one sample line
Sample Line too long
I.D. too small
Poor gate seal
Clear restriction
Replace filter
Reduce number of
analyzers on same Line
Correct sample line issue
Contact Honeywell
Analytics Service
102
LOW CC
WARNING
Chemcassette
Low
Days
Remaining
Chemcassette counter
decremented to less than
120 windows
Less than one
day remains on
Chemcassette
Replace Chemcassette
Summary of Contents for Vertex M
Page 8: ...1 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 1 Introduction ...
Page 25: ...Vertex MTM 24 Point Continuous Monitor Section 1 Introduction 1 18 ...
Page 26: ...2 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 2 Installation ...
Page 35: ...Vertex MTM 24 Point Continuous Monitor Section 2 Installation 2 10 ...
Page 36: ...3 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 3 Startup ...
Page 40: ...Vertex MTM 24 Point Continuous Monitor Section 3 Startup 3 5 Figure 3 1 Vertex M Main Screen ...
Page 77: ...4 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 4 Operation ...
Page 93: ...Vertex MTM 24 Point Continuous Monitor Section 4 Operation 4 17 Figure 4 9 Filter Options ...
Page 109: ...Vertex MTM 24 Point Continuous Monitor Section 4 Operation 4 33 ...
Page 125: ...5 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 5 Maintenance ...
Page 134: ...Vertex MTM 24 Point Continuous Monitor Section 5 Maintenance 5 10 ...
Page 140: ...Vertex MTM 24 Point Continuous Monitor Section 5 Maintenance 5 16 ...
Page 141: ...6 1 Vertex M TM Technical Handbook 6 Troubleshooting ...
Page 164: ...Section 6 Troubleshooting 6 24 ...
Page 166: ...Vertex MTM 24 Point Continuous Monitor Appendix A Installation Drawings A 2 A 1 Introduction ...
Page 170: ...Vertex MTM 24 Point Continuous Monitor Appendix A Installation Drawings A 6 ...
Page 173: ...Vertex MTM 24 Point Continuous Monitor Appendix A Installation Drawings A 9 ...
Page 174: ...B 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook A Specifications ...
Page 178: ...C 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook A Detectable Gases ...
Page 193: ...Vertex MTM 24 Point Continuous Monitor Appendix A Replacement and Consumable Items D 4 ...
Page 240: ...Vertex MTM 24 Point Continuous Monitor Appendix A 4 20mA Analog Output Option G 4 ...
Page 250: ...Vertex MTM 24 Point Continuous Monitor Appendix A Warranty Statement I 4 ...