Vertex M
TM
24-Point Continuous Monitor
Section 4 - Operation
4-38
Auto Balance the Flow Rate
Note:
All analyzers should be in Monitor except for the one
being Auto Balanced.
Perform an Auto Balance for the following reasons:
• When a line length is changed
• If end-of-line filters are replaced
• The gas family changes
• When a new Chemcassette
®
is installed
• When analyzers are added to a Vertex M rack
To perform an Auto Balance:
1. Make sure Chemcassette
®
is installed.
2. Set analyzer to idle (not monitoring). See
Section 4.6.1 Run Time Options
for the
procedure.
3. If not already open, touch Main Screen, Menu,
Calibration. The Flow Diagnostic window
opens.
4. Choose analyzer from the selection pad in the
upper right-hand corner.
5. Touch the Auto Balance function button. Vertex
M will characterize the flow between minimum
and maximum. This data will be used to set
the flow to 180 cc/min. when you return the
analyzer to monitor mode. Autobalancing takes
approximately 130 seconds to complete.
6. If required, repeat for other analyzers.
Note:
If Vertex M cannot adjust the flow to the proper range
(150 to 210 cc per minute at worst case condition)
consult Honeywell Analytics service.
Summary of Contents for Vertex M
Page 8: ...1 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 1 Introduction ...
Page 25: ...Vertex MTM 24 Point Continuous Monitor Section 1 Introduction 1 18 ...
Page 26: ...2 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 2 Installation ...
Page 35: ...Vertex MTM 24 Point Continuous Monitor Section 2 Installation 2 10 ...
Page 36: ...3 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 3 Startup ...
Page 40: ...Vertex MTM 24 Point Continuous Monitor Section 3 Startup 3 5 Figure 3 1 Vertex M Main Screen ...
Page 77: ...4 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 4 Operation ...
Page 93: ...Vertex MTM 24 Point Continuous Monitor Section 4 Operation 4 17 Figure 4 9 Filter Options ...
Page 109: ...Vertex MTM 24 Point Continuous Monitor Section 4 Operation 4 33 ...
Page 125: ...5 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 5 Maintenance ...
Page 134: ...Vertex MTM 24 Point Continuous Monitor Section 5 Maintenance 5 10 ...
Page 140: ...Vertex MTM 24 Point Continuous Monitor Section 5 Maintenance 5 16 ...
Page 141: ...6 1 Vertex M TM Technical Handbook 6 Troubleshooting ...
Page 164: ...Section 6 Troubleshooting 6 24 ...
Page 166: ...Vertex MTM 24 Point Continuous Monitor Appendix A Installation Drawings A 2 A 1 Introduction ...
Page 170: ...Vertex MTM 24 Point Continuous Monitor Appendix A Installation Drawings A 6 ...
Page 173: ...Vertex MTM 24 Point Continuous Monitor Appendix A Installation Drawings A 9 ...
Page 174: ...B 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook A Specifications ...
Page 178: ...C 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook A Detectable Gases ...
Page 193: ...Vertex MTM 24 Point Continuous Monitor Appendix A Replacement and Consumable Items D 4 ...
Page 240: ...Vertex MTM 24 Point Continuous Monitor Appendix A 4 20mA Analog Output Option G 4 ...
Page 250: ...Vertex MTM 24 Point Continuous Monitor Appendix A Warranty Statement I 4 ...