Section 6 - Troubleshooting
6-7
Event
Number
Event List Desc.
(Short)
Event History
Description
Data Field
Problem
Possible Cause
Recovery
111
DACS COM FAIL No
Communication
from PC To Az
0.0
Ethernet communica-
tions
Communications to
DACS interrupted longer
than 20 seconds
Check Ethernet cable at
rear of analyzer
Check Ethernet hub
connection and operation
Check Ethernet
connection to DACS
Restart DACS to clear
OPC driver problem
112
SWAPPED BAD
PUMP
Single pump
failed and
swapped
Number of
bad pump - 1
or 2
Pump swap has occurred
because vacuum levels
fell below 6 in. Hg
Pump failure
Rebuild/Replace non-
operating pump
Unused slots not plugged Install 1295-0404
connector plug on unused
slot
Large 0.018 in transport
flow orifices installed
Replace all orifices with
0.015 in orifices (p/n
1290K0009 per Analyzer)
Check valve on inactive
pump leaking
Replace check valve
Contact Honeywell
Analytics Service
113
OPTICS MAINT
Optics Drive
High-Cleaning
Req
Drive
LED drive required to set
zero too high
Clean optics
Contact Honeywell
Analytics Service
Tape leader installed
improperly
Leader not centered (Pts
1 or 5 error)
Reload Chemcassette and
recalibrate using leader
Autozero performed while
tape on light or dark gray
portion of leader
Reload Chemcassette
and recalibrate using
leader
Dirt in optics
Clean optics block
114
POINT VAC ERR Excessive Point
Vacuum
in. Hg
Sample vacuum is
greater than 5 in. Hg
vacuum for longer than 5
seconds
End of line filter clogged
Replace clogged end of
line filter
Sample line kinked
Isolate by disconnecting
possible crimped sample
line
Maximum line length
exceeded on a shared
sample line
Isolate shared sample line
Review limits number of
analyzers, tubing ID vs.
Length (See
Table B.2
)
Crimped tube in cable
carrier
Identify crimps in tubing
harness by checking
pressure with analyzer
closed vs. open
Summary of Contents for Vertex M
Page 8: ...1 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 1 Introduction ...
Page 25: ...Vertex MTM 24 Point Continuous Monitor Section 1 Introduction 1 18 ...
Page 26: ...2 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 2 Installation ...
Page 35: ...Vertex MTM 24 Point Continuous Monitor Section 2 Installation 2 10 ...
Page 36: ...3 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 3 Startup ...
Page 40: ...Vertex MTM 24 Point Continuous Monitor Section 3 Startup 3 5 Figure 3 1 Vertex M Main Screen ...
Page 77: ...4 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 4 Operation ...
Page 93: ...Vertex MTM 24 Point Continuous Monitor Section 4 Operation 4 17 Figure 4 9 Filter Options ...
Page 109: ...Vertex MTM 24 Point Continuous Monitor Section 4 Operation 4 33 ...
Page 125: ...5 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 5 Maintenance ...
Page 134: ...Vertex MTM 24 Point Continuous Monitor Section 5 Maintenance 5 10 ...
Page 140: ...Vertex MTM 24 Point Continuous Monitor Section 5 Maintenance 5 16 ...
Page 141: ...6 1 Vertex M TM Technical Handbook 6 Troubleshooting ...
Page 164: ...Section 6 Troubleshooting 6 24 ...
Page 166: ...Vertex MTM 24 Point Continuous Monitor Appendix A Installation Drawings A 2 A 1 Introduction ...
Page 170: ...Vertex MTM 24 Point Continuous Monitor Appendix A Installation Drawings A 6 ...
Page 173: ...Vertex MTM 24 Point Continuous Monitor Appendix A Installation Drawings A 9 ...
Page 174: ...B 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook A Specifications ...
Page 178: ...C 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook A Detectable Gases ...
Page 193: ...Vertex MTM 24 Point Continuous Monitor Appendix A Replacement and Consumable Items D 4 ...
Page 240: ...Vertex MTM 24 Point Continuous Monitor Appendix A 4 20mA Analog Output Option G 4 ...
Page 250: ...Vertex MTM 24 Point Continuous Monitor Appendix A Warranty Statement I 4 ...