Section 6 - Troubleshooting
6-10
Event
Number
Event List Desc.
(Short)
Event History
Description
Data Field
Problem
Possible Cause
Recovery
127
AUTOBALANCE
FAIL
Flow
Autobalance
Failed
No. of Failed
Attempts
Detailed failures can be determined by the data from the accompanying info
event(s) by using the table below.
11 to 23
Adjusting factor out of
range
Anomaly encountered
Repeat Auto balance
Faulty valve
Contact HA Service
30
Adjusting factor varies
between 3 spots
Anomaly encountered
Repeat Auto balance
Other
Contact HA Service
41 to 43
Inadequate flow
Insufficient vacuum
(<8 in Hg)
See recovery for Fault
112
Excessive pressure drop
in tubing
See recovery for Fault
114
Faulty valve
Contact HA Service
51 to 53
Excessive minimum flow
Debris in valve
Repeat Auto balance
Faulty valve
Contact HA Service
128*
WINDOW ZERO
RESET
Window Zero
Reset
Diagnostic
Counter
Same as Fault 126
Same as Fault 126
Same as Fault 126
129
ACCEL. CC
USAGE
Acclerated
Chemcassette
Usage
Time
since last
Chemcassette
advance
Chemcassette advancing
more often that expected
Low level background
gas below lower
detectable limit
Locate source of
background gas
130
VR FAILED
Virtual
Reference Test
Failed
Sample
Counter
Tape reflectance differs
slightly between windows
Old tape
Contact HA Service if this
occurs frequently
Ambient gas
concentration
*Dependent on the setting in the Configuration menu.
Summary of Contents for Vertex M
Page 8: ...1 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 1 Introduction ...
Page 25: ...Vertex MTM 24 Point Continuous Monitor Section 1 Introduction 1 18 ...
Page 26: ...2 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 2 Installation ...
Page 35: ...Vertex MTM 24 Point Continuous Monitor Section 2 Installation 2 10 ...
Page 36: ...3 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 3 Startup ...
Page 40: ...Vertex MTM 24 Point Continuous Monitor Section 3 Startup 3 5 Figure 3 1 Vertex M Main Screen ...
Page 77: ...4 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 4 Operation ...
Page 93: ...Vertex MTM 24 Point Continuous Monitor Section 4 Operation 4 17 Figure 4 9 Filter Options ...
Page 109: ...Vertex MTM 24 Point Continuous Monitor Section 4 Operation 4 33 ...
Page 125: ...5 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 5 Maintenance ...
Page 134: ...Vertex MTM 24 Point Continuous Monitor Section 5 Maintenance 5 10 ...
Page 140: ...Vertex MTM 24 Point Continuous Monitor Section 5 Maintenance 5 16 ...
Page 141: ...6 1 Vertex M TM Technical Handbook 6 Troubleshooting ...
Page 164: ...Section 6 Troubleshooting 6 24 ...
Page 166: ...Vertex MTM 24 Point Continuous Monitor Appendix A Installation Drawings A 2 A 1 Introduction ...
Page 170: ...Vertex MTM 24 Point Continuous Monitor Appendix A Installation Drawings A 6 ...
Page 173: ...Vertex MTM 24 Point Continuous Monitor Appendix A Installation Drawings A 9 ...
Page 174: ...B 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook A Specifications ...
Page 178: ...C 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook A Detectable Gases ...
Page 193: ...Vertex MTM 24 Point Continuous Monitor Appendix A Replacement and Consumable Items D 4 ...
Page 240: ...Vertex MTM 24 Point Continuous Monitor Appendix A 4 20mA Analog Output Option G 4 ...
Page 250: ...Vertex MTM 24 Point Continuous Monitor Appendix A Warranty Statement I 4 ...