Section 6 - Troubleshooting
6-6
Event
Number
Event List Desc.
(Short)
Event History
Description
Data Field
Problem
Possible Cause
Recovery
103
TIMEOUT IDLE
Timeout Of
Monitor
0.0
Analyzer in IDLE, LOAD
CC, FLOW
Out of Monitor time limit
has been exceeded
Press RESET to
reinitialize timer.
User error
Enter Monitor using
Runtime Options
Return analyzer to
Monitor Mode
Time limit too short
Change time limit in
Configuration Profile
104
TIMEOUT PNT
DIS
Timeout Run-
time Point
Disable
0.0
Operator error
Point disabled longer
than time limit setting in
the Configuration Profile
Press RESET to
reinitialize timer
Restore point to Monitor
Mode
Time limit too short
Change time limit in
Configuration Profile
105
TIMEOUT ALM
DIS
Timeout Run-
time Alarm
Disable
0.0
Operator error
Point disabled longer
than time limit setting in
the Configuration Profile
Press RESET to
reinitialize timer
Restore point to Monitor
Mode
Time limit too short
Change time limit in
Configuration Profile
106
POS PNT
PRESSURE
Positive Point
Pressure
Pressure in.
Hg
Positive pressure
between sampling
location and monitor
Point pressure above
atmospheric pressure
while in idle
Relieve/reduce pressure
Transducer error
Calibration error
Contact Honeywell
Analytics Service
Defective sensor
Contact Honeywell
Analytics Service
Offset incorrect
Contact Honeywell
Analytics Service
108
FLOW ADJ ERR
HI
High Flow After
Adjustment
cc/min.
Flow cannot be adjusted
below 210 cc/min.
Debris in proportional
valve
Verify range of valve
Perform Auto Balance to
clear debris
Debris or defective valve
Proportional valve held
open
Replace with
0235-1279
109
CC EXPIRED
Chemcassette
Passed
Expiration Date
0.0
Operator error
Chemcassette installed
past its expiration date
Replace Chemcassette
Chemcassette expired
Expiration date reached
Replace Chemcassette
Summary of Contents for Vertex M
Page 8: ...1 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 1 Introduction ...
Page 25: ...Vertex MTM 24 Point Continuous Monitor Section 1 Introduction 1 18 ...
Page 26: ...2 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 2 Installation ...
Page 35: ...Vertex MTM 24 Point Continuous Monitor Section 2 Installation 2 10 ...
Page 36: ...3 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 3 Startup ...
Page 40: ...Vertex MTM 24 Point Continuous Monitor Section 3 Startup 3 5 Figure 3 1 Vertex M Main Screen ...
Page 77: ...4 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 4 Operation ...
Page 93: ...Vertex MTM 24 Point Continuous Monitor Section 4 Operation 4 17 Figure 4 9 Filter Options ...
Page 109: ...Vertex MTM 24 Point Continuous Monitor Section 4 Operation 4 33 ...
Page 125: ...5 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 5 Maintenance ...
Page 134: ...Vertex MTM 24 Point Continuous Monitor Section 5 Maintenance 5 10 ...
Page 140: ...Vertex MTM 24 Point Continuous Monitor Section 5 Maintenance 5 16 ...
Page 141: ...6 1 Vertex M TM Technical Handbook 6 Troubleshooting ...
Page 164: ...Section 6 Troubleshooting 6 24 ...
Page 166: ...Vertex MTM 24 Point Continuous Monitor Appendix A Installation Drawings A 2 A 1 Introduction ...
Page 170: ...Vertex MTM 24 Point Continuous Monitor Appendix A Installation Drawings A 6 ...
Page 173: ...Vertex MTM 24 Point Continuous Monitor Appendix A Installation Drawings A 9 ...
Page 174: ...B 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook A Specifications ...
Page 178: ...C 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook A Detectable Gases ...
Page 193: ...Vertex MTM 24 Point Continuous Monitor Appendix A Replacement and Consumable Items D 4 ...
Page 240: ...Vertex MTM 24 Point Continuous Monitor Appendix A 4 20mA Analog Output Option G 4 ...
Page 250: ...Vertex MTM 24 Point Continuous Monitor Appendix A Warranty Statement I 4 ...