Faults
91
Message
Description, process
I1 Warn. sig.
I2 Warn. sig.
I3 Warn. sig.
I4 Warn. sig.
When an input is configured as a warning input, if an
active signal is present on the relevant terminal (i.e.
terminal DI2) this message is displayed.
The input configurations can be changed using the pa-
rameterization program "enSuite".
Batt. low
The remaining battery service life is less than 6 months.
The batteries must be replaced soon.
c) Remarks:
Clock n. set
Calibration of the volume corrector's internal clock has
not yet been carried out by the factory.
Encoder err.
On the encoder connected to input 1 or input 2 no meter
reading could be detected for 20 seconds (with external
voltage supply) or for the duration of an encoder readout
cycle
“
EZyk
” (with battery power supply).
CDL full
The Certification Data Log is full. A change to "PL"
identified access parameters without calibration lock is
not possible, until the contents of the Certification Data
Log are cleared.
U.Logb. full
The software Update Log is full. An update is not possi-
ble until the contents of the software update log are
cleared.
online
Data is currently being transmitted via an interface
(modem or optical interface).
If this message is displayed, the "o" will flash in the "Sta-
tus" field of the display (see chapter 6.3.4, page 58).
I1-Rep. sig.
I2-Rep. sig.
I3-Rep. sig.
I4-Rep. sig.
When an input is configured as a remark input, if an
active signal is present on the relevant terminal (i.e.
terminal DI2) this message is displayed.
The input configurations can be changed using the
parameterization program "enSuite".
Summary of Contents for DL230
Page 6: ......
Page 98: ...98 Appendix 9 3 EC Declaration of Conformity ...
Page 99: ...Appendix 99 9 4 ATEX Type Examination Certificate 9 4 1 Associated apparatus for Zone 0 1 ...
Page 100: ...100 Appendix ...
Page 101: ...Appendix 101 ...
Page 102: ...102 Appendix ...
Page 103: ...Appendix 103 ...
Page 104: ...104 Appendix ...