Maintenance
Ill.: 22
9. Fix the new battery into place with
the mounting element. The mounting
element can then be fixedly pushed
into the left- and right-hand guides.
10. Remove the old battery from the
housing and dispose of it correctly.
Close the housing using the screws
provided (Ill.: 20 2 ).
12. Re-enter the start capacity of the
batteries, as described in chapter
7.2.2.
13. Check the calculated service life as described in chapter 7.2.3. At
standard measurement conditions (see chapter 3.2.1), the service life
should display at least 96 months. If not, compare the settings with
those for standard measurement conditions and repeat step 12 if nec-
essary.
Please ensure that the new batteries are connected
correctly and are in a fixed position inside the DL230.
CAUTION!
Damage to property caused by improper device
closure!
If the device is not properly closed cables can get pinched
and cause damage.
Therefore:
– Ensure correct positioning of the cable guide when
closing.