90
Faults
Message
Description, process
b) Warnings:
Data restore
DL230 batteries failed intermittently. As a result of this,
the time has not changed and measurement acquisition
has failed. All data has been saved.
This message appears when the battery is changed and
the old batteries are removed before the new ones are
connected. (see chapter 7.2, page 76).
Process using parameterization program "enSuite":
◼
Set the DL230 clock
◼
Clear the status register
Outp.1 Error
Outp.2 Error
More pulses need to be sent to the named output, than
its programming permits.
To eliminate the cause of this problem, use the "enSuite"
parameterization program to
◼
either reduce the cp value of the output
◼
or increase the output frequency.
In both cases, the change must be aligned with the de-
vice connected to the DL230.
I1-Warning
I2-Warning
The encoder could not be read at the end of the meas-
urement period. The flow recording at this point is invalid.
Update error
An error in the cached software image was found prior to
the software update.
Softw. error
This message is used for factory diagnostics.
Please contact the Elster customer service team (see
chapter 1.3 "Customer service").
Sett. error
Due to the programming carried out an unusable settings
combination has been created.
Please contact the Elster customer service team
(see chapter 1.3 "Customer service").
Summary of Contents for DL230
Page 6: ......
Page 98: ...98 Appendix 9 3 EC Declaration of Conformity ...
Page 99: ...Appendix 99 9 4 ATEX Type Examination Certificate 9 4 1 Associated apparatus for Zone 0 1 ...
Page 100: ...100 Appendix ...
Page 101: ...Appendix 101 ...
Page 102: ...102 Appendix ...
Page 103: ...Appendix 103 ...
Page 104: ...104 Appendix ...