140
Chapter 6: Device operation
2. Navigate to the home screen by turning the MFK to "System":
3. Press the MFK.
4. Navigate to "Tightness Check" by turning the MFK.
5. Press the MFK
22:33
60
PCV
Datum und Uhrzeit
13.02.16
Anzahl der Beatmungssets
3
Anwendermodus Klinik
Dichtigkeitstest
Beatmungsstd. seit Reset
65
Beatmungsstd. Gesamt
342
Betriebsstunden 622
Gebläse-Service in
14658
SW-Version 2.002
Seriennummer EAG1300001
10:36
h
h
h
h
Läuft...
Test beenden
22,7
p
0
Flow
[ l/min]
[ hPa]
The pressure is gradually inceased to the maximum pressure by the device.
Observe the flow value. If the flow value exceeds 5 l/min during the tightness
check, there is a leak in the tube circuit. In that case, check the entire circuit
for leaks in tubes and at tube connections. As long as the flow remains under
5 l/min the tightness check is positive. Press the MFK to complete the test.
If the pressure (p) exceed 60 hPa during the tightness check, the device
must be returned for servicing.