198
Chapter 10: Appendix
MANUFACTURER’S DECLARATION ON ELECTRO
-
MAGNETIC COMPATIBILITY
CARAT II pro satisfies the IEC 60601-1-2:2014 standard and is intended for use
in the electromagnetic environment described below. Deviating ambient condi-
tions can impair the essential performance features such as pressure accuracy
and alarm, or lead to the breakdown of the device.
Guidance and manufacturer’s declaration – electromagnetic emissions
The CARAT II pro ventilator
is intended for use in the electromagnetic environment spec-
ified below. The user of the
CARAT II pro ventilator
should ensure that it is used in such
an environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions acc. to
CISPR 11
Group 1
The
CARAT II pro
ventilator uses RF energy
only for its internal function. Therefore, RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
RF emissions acc. to
CISPR 11
Class B
The
CARAT II pro
ventilator is suitable for use
in all establishments including those directly
connected to the public low-voltage power
supply network that supplies buildings used
for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage
fluctuations/flicker emis
-
sions acc. to IEC 61000-
3-3
complies
Summary of Contents for CARAT II pro
Page 2: ......
Page 22: ...22 Chapter 1 Introduction ...
Page 40: ...40 Chapter 3 Description of device ...
Page 106: ...106 Chapter 5 Ventilation modes ...
Page 148: ...148 Chapter 6 Device operation ...
Page 168: ...168 Chapter 8 Cleaning and disinfection ...
Page 175: ...Chapter 10 Appendix 175 Chapter 10 APPENDIX ...