![Hoffrichter CARAT II pro User Manual Download Page 52](http://html.mh-extra.com/html/hoffrichter/carat-ii-pro/carat-ii-pro_user-manual_2144201052.webp)
52
Chapter 4: Commissioning
CONNECTING A HUMIDIFIER
A humidifier is used to humidify and warm the breathing air. If you use a humidi
-
fier, be sure to follow the manufacturer's user's manual.
The humidifier is integrated into the inspiration section. It should be positioned
below the patient and the device, so that no water can accumulate in the patient's
lung or in the device.
Note: When using a bacterial filter, please read the instructions of use of this filter.
Single line patient circuit
Connect the tube circuit to the humidifier and the device as shown in Figure 18.
When using a tube circuit without pressure measuring tube, the used tube circuit
must be calibrated page 55).
I
PAT
IEN
T
C
A
B
H
J
F
E
D
G
A
Patient side connection
B
Expiration valve
C
Ventilation tube
D
Connecting tube for inspiration
E
Bacterial filter
F
Adapter for bacterial filter
G
Control tube
H
Pressure measuring tube
I
Humidifier
J
Air outlet
Figure 18:
Connecting the humidifier - single line patient circuit