Chapter 4: Commissioning
3. Navigate to the system screen using the MFK
. Press the MFK.
4. Navigate to "Calibrate FiO
2
Sensor" by turning the MFK.
22:33
60
PCV
Calibrate Tube
Calibrate O
2
Sensor
FiO
2
-Monitoring Internal
Alarm Volume
3
Night Screen
Meaurements
Timer Screen Change
2
Brightness Display
20
Brightness LEDs
20
Brightness MFK 20
Language English
min
%
%
%
5. Press the MFK. Calibration begins.
6. If the calibration was successful, "OK" will appear after a few seconds. If the cali-
bration was not successful, "Error" will display. In the event of an error, repeat
the calibration. If the calbration is still unsuccessful, replace the FiO
2
sensor.
22:33
60
PCV
Calibrate Tube
Calibrate O
2
Sensor
FiO
2
-Monitoring Internal
Alarm Volume
3
Night Screen
Meaurements
Timer Screen Change
2
Brightness Display
20
Brightness LEDs
20
Brightness MFK 20
Language English
min
%
%
%
21% Ok
Finish calibration
7. Press the MFK to end the FiO
2
sensor calibration.
Depending on environmental conditions and the storage time, the sensor may take
up to 15 minutes after connecting to reach signal stability again.
FiO
2
sensors
have a limited service life. The service life of the sensors is
approx. 1 year at a oxygen concentration for about 40 %. After that, the
FiO
2
sensor
must be replaced by a new one. The sensor should not be storaged
more than 6 month. For the longest possible sensor service life, we recom-
mend storage at +5 °C to + 30 °C.
Summary of Contents for CARAT II pro
Page 2: ......
Page 22: ...22 Chapter 1 Introduction ...
Page 40: ...40 Chapter 3 Description of device ...
Page 106: ...106 Chapter 5 Ventilation modes ...
Page 148: ...148 Chapter 6 Device operation ...
Page 168: ...168 Chapter 8 Cleaning and disinfection ...
Page 175: ...Chapter 10 Appendix 175 Chapter 10 APPENDIX ...