![Hoffrichter CARAT II pro User Manual Download Page 65](http://html.mh-extra.com/html/hoffrichter/carat-ii-pro/carat-ii-pro_user-manual_2144201065.webp)
Chapter 4: Commissioning
65
USING OXYGEN
Before using oxygen, the safety instructions must be read as of page 29.
Oxygen may only be supplied during active ventilation.
The supply of oxygen is possible in all ventilation modes. Please note that any
changes to the ventilation parameters, as e.g. pressure, I:E, frequency, will lead
to a change of the FiO
2
concentration.
CONNECTING THE OXYGEN SOURCE
Only the oxygen connection adapter supplied may be used to connect oxygen.
Otherwise, there is a risk that the back-stop in the connection is damaged.
Connect the oxygen source to the device as shown in Figure 28.
0
I
C
B
A
A
Oxygen connection
B
Oxygen connection adapter, straight
C
Tube from the oxygen source
Figure 28:
Connecting the oxygen source (rear of unit)