Chapter 5: Ventilation modes
105
HIGH SpO
2
PCV
APCV
PSV
PSV-S
VCV
AVCV
P-SIMV
V-SIMV
CPAP
If the measured oxygen saturation is greater than the "High SpO
2
" setting, then
the "High SpO
2
" alarm will be triggered.
LOW SpO
2
PCV
APCV
PSV
PSV-S
VCV
AVCV
P-SIMV
V-SIMV
CPAP
If the measured oxygen saturation is lower than the "Low SpO
2
" setting, the "SpO
2
too low" alarm will be triggered.
HIGH HEART RATE
PCV
APCV
PSV
PSV-S
VCV
AVCV
P-SIMV
V-SIMV
CPAP
If the measured pulse frequency is higher than the "High Heart Rate" setting, the
"High Heart Rate" alarm will be triggered.
LOW HEART RATE
PCV
APCV
PSV
PSV-S
VCV
AVCV
P-SIMV
V-SIMV
CPAP
If the measured pulse frequency is lower than the "Low Heart Rate" setting, the
"Low Heart Rate" alarm will be triggered.
HIGH PRESSURE TOLERANCE
PCV
APCV
PSV
PSV-S
VCV
AVCV
P-SIMV
V-SIMV
CPAP
If the measured pressure is higher than the set pressure plus the setting
"High Pressure Tolerance", then the alarm "High Pressure" will be triggered.
LOW PRESSURE TOLERANCE
PCV
APCV
PSV
PSV-S
VCV
AVCV
P-SIMV
V-SIMV
CPAP
If the measured pressure is lower than the set pressure minus the setting "Low
Pressure Tolerance" then the "Low Pressure" alarm will be triggered.
Summary of Contents for CARAT II pro
Page 2: ......
Page 22: ...22 Chapter 1 Introduction ...
Page 40: ...40 Chapter 3 Description of device ...
Page 106: ...106 Chapter 5 Ventilation modes ...
Page 148: ...148 Chapter 6 Device operation ...
Page 168: ...168 Chapter 8 Cleaning and disinfection ...
Page 175: ...Chapter 10 Appendix 175 Chapter 10 APPENDIX ...