180
Chapter 10: Appendix
ERROR MESSAGES
Table 13:
Error messages during operation and at device start-up
Error message
Cause
Correction
Error SpO
2
sensor
Communication to SpO
2
measuring module not
possible
Device must be serviced
SpO
2
sensor defective
Replace SpO
2
sensor
SD card is full
No storage space avail-
able on the SD card
Insert blank SD card
Flash Not Working
No access to the flash
Device must be serviced
Default Parameters
Loaded
No valid parameter
set available or they are
faulty
see page 181
Error Flow Sensor
Flow sensor is defective
Device must be serviced
Error FiO
2
Sensor
FiO
2
sensor is defective
Check FiO
2
sensor
connection
Recalibrate FiO
2
sensor
Replace FiO
2
sensor
Error Pressure Sensor
Defective pressure sensor
Device must be serviced
Calibration File Damaged
Calibration data for
sensors is damaged
Device must be serviced
Eventlog File Damaged
Event read data failed
Device must be serviced
Primary Alarm Not Working
Primary alarm sound unit
is defective
Device must be serviced
Secondary
Alarm Not Working
Secondary alarm sound
transmitter is defective
Device must be serviced
No Alarm Available
Primary and second-
ary alarm sound units are
defective
Device must be serviced
Booting Error
Boot failed
Device must be serviced
Fatal Error
Fatal error occurred
Device must be serviced
Maintenance 5 necessary
Maximum blower run times
reached
Device must be serviced
Summary of Contents for CARAT II pro
Page 2: ......
Page 22: ...22 Chapter 1 Introduction ...
Page 40: ...40 Chapter 3 Description of device ...
Page 106: ...106 Chapter 5 Ventilation modes ...
Page 148: ...148 Chapter 6 Device operation ...
Page 168: ...168 Chapter 8 Cleaning and disinfection ...
Page 175: ...Chapter 10 Appendix 175 Chapter 10 APPENDIX ...