Chapter 10: Appendix
INSTRUCTIONS FOR VENTILATION OUTSIDE OF A
CLINICAL SETTING
You should take the following into consideration before discharging a patient requi-
ring at home ventilation.
1. Limit the number of sets needed by the patient should use.
2. Adjust all parameters and settings for home ventilation.
3.
Finally, switch the user profile to “Home”.
4. Train the patient in the use of this device. In addition, explain which accessory
equipment may be used and how to maintain and care for the device (e.g.
battery care, cleaning the device).
5. Discuss how the patient should act, particularly if he or she is in danger or
if an alarm sounds. Provide a phone number to the patient to call in case of
emergency.
TECHNICAL SPECIFICATIONS FOR THE DEVICE
ESSENTIAL PERFORMANCE FEATURES
The essential performance features of the CARAT II pro are:
• pressure-controlled mechanical support with a pressure accuracy of
±
(2 % of the scale value + 8 % of the actual measured value),
• sounding of an alarm when inspiration pressure is exceeded, and alert by
remote alarm,
• the volume-controlled, mechanical support with a volume accuracy of
±
20 %
for a tidal volume above 100 ml or at a minute volume above 3 l/min,
• sounding of an alarm when the alarm parameter “Low Inspiration Volume” is
not met or when the alarm parameter “High Inspiration Volume” is exceeded,
and alert by remote alarm,
• sounding of an alarm when the alarm parameter “Low Expiration Volume” is
not met or when the alarm parameter “High Expiration Volume” is exceeded,
and alert by remote alarm,
• sounding of an alarm when the alarm parameter “Low Frequency” is not met or
when the alarm parameter “High Frequency” is exceeded, and alert by remote
alarm,
• sounding of an alarm when the alarm parameter “Low FiO
2
” is not met and alert
by remote alarm (only with an external FiO
2
monitoring),
• the alarm “Internal Battery Empty” in case of imminent failure of the mains supply
(< 1 min) and the distribution by remote alarm and
Summary of Contents for CARAT II pro
Page 2: ......
Page 22: ...22 Chapter 1 Introduction ...
Page 40: ...40 Chapter 3 Description of device ...
Page 106: ...106 Chapter 5 Ventilation modes ...
Page 148: ...148 Chapter 6 Device operation ...
Page 168: ...168 Chapter 8 Cleaning and disinfection ...
Page 175: ...Chapter 10 Appendix 175 Chapter 10 APPENDIX ...