Chapter 5: Ventilation modes
101
Using SMART Function
The inspiration trigger is
not
initiated by artefacts.
Δ
t
Δ
T
t (s)
V (ml)
Volume curve
Trigger activity
t (s)
Trigger threshold
Triggering
Suppression
false triggering
Trigger threshold
V (ml)
Artefacts Spontaneous breathing
Trigger by patient (spontaneous breathing)
Figure 45:
Inspiration trigger using SMART Function
If the "SMART Function" is active, the device distinguishes between patient efforts
and artefacts.
An inspiration trigger is initiated according to the following conditions:
A time period
∆
T is calculated depending on the IPAP and starting with the expi-
ration period. If the inspiration trigger threshold is reached, an additional IPAP-
dependent time period
∆
t is calculated. If a patient effort meets the trigger con-
dition within this time period
∆
t, an inspiration trigger will initiate. If this is not the
case, the device supposes an artefact and an inspiration trigger will not initiate.
After the time period
∆
T has expired, an inspiration trigger is immediately initiated
after exceeding the trigger thresholds.
If an auto trigger is occurred although the SMART Function is active, we
recommend to increase the trigger level or to set a trigger lock.
Summary of Contents for CARAT II pro
Page 2: ......
Page 22: ...22 Chapter 1 Introduction ...
Page 40: ...40 Chapter 3 Description of device ...
Page 106: ...106 Chapter 5 Ventilation modes ...
Page 148: ...148 Chapter 6 Device operation ...
Page 168: ...168 Chapter 8 Cleaning and disinfection ...
Page 175: ...Chapter 10 Appendix 175 Chapter 10 APPENDIX ...