![Hoffrichter CARAT II pro User Manual Download Page 159](http://html.mh-extra.com/html/hoffrichter/carat-ii-pro/carat-ii-pro_user-manual_2144201159.webp)
Chapter 7: Alarms and messages
159
Table 8:
Fixed Alarms
Alarm
Priority
Audible
alarm
State of the
alarm LED
Cause
Correction
Error FiO
2
sensor
1
MEDIUM C b a
Yellow -
flashes
FiO
2
sensor
disconnected
from the decice
Connect the FiO
2
sensor to the
device
Check
SpO
2
sensor
2
MEDIUM C b a
Yellow -
flashes
SpO
2
sensor
disconnected
from the decice
Connect the
SpO
2
sensor
to the device
SpO
2
sensor
disconnected
from the patient
Connect the
SpO
2
sensor
to the patient
1 The alarm message is only possible, if at least one of alarms “High FiO
2
“ or “Low FiO
2
”
is enabled.
2 The alarm message is only possible, if at least one of alarms “High SpO
2
”, “Low SpO
2
”,
“High Heart Rate“ oder “Low Heart Rate” is enabled.