152
Chapter 7: Alarms and messages
AUDIO ALARMS VOLUME
The audio alarm volume can be set to 3 levels in the system screen:
• Level 1
→
low volume (55 dB)
• Level 2
→
medium volume (60 dB)
• Level 3
→
high volume (65 dB)
The volume of levels 1 and 2 are automatically raised to a level 3 in case the alarm
event still exists after 1 minute of audible alarm.
VISIBLE ALARM OUTPUT
Visible alarms are displayed as follows:
• via the alarm LED
• in the toolbar
• as a textbox
• lighting up the multifunctional knob
ALARM OUTPUT VIA THE ALARM LED
The alarm LED may take on 3 different statuses, to signify the current alarm priority.
•
Red, flashes rapidly (2 Hz)
→
high-priority alarm
•
Yellow, flashes (0.5 Hz)
→
medium-priority alarm
• Turquoise, glows steadily
→
low-priority alarm
If multiple alarms are triggered simultaneously or in quick succession, the alarm
with the highest priority will be displayed first.
More information on alarm LEDs is available on page 37.
ALARM OUTPUT IN THE TOOLBAR
Alarms are shown in the toolbar by the "Alarm active" icon and displayed with an
alarm message. The icon's color indicates the alarm priority:
• Red icon
→
high-priority alarm
• Yellow icon
→
medium-priority alarm
• Turquoise icon
→
low-priority alarm
If multiple alarms are triggered simultaneously or in quick succession, the alarm
with the highest priority will be displayed first.
Summary of Contents for CARAT II pro
Page 2: ......
Page 22: ...22 Chapter 1 Introduction ...
Page 40: ...40 Chapter 3 Description of device ...
Page 106: ...106 Chapter 5 Ventilation modes ...
Page 148: ...148 Chapter 6 Device operation ...
Page 168: ...168 Chapter 8 Cleaning and disinfection ...
Page 175: ...Chapter 10 Appendix 175 Chapter 10 APPENDIX ...