Chapter 4: Commissioning
43
2. Connect the mains cable to the power supply.
3. Insert the mains cable plug into the power socket (100 - 240 V, 50/60 Hz).
0
I
D
E
B
A
C
A
Power socket
B
Mains cable
C
Power supply
D
Power supply plug
E
DC connector socket
Figure 9:
Mains connection via power supply unit
4. The device boots and performs the following hardware tests:
• Testing the primary and secondary alarm sounds: Both alarm sounds give
a short beep one-by-one.
• Checking of other hardware components.
If errors are detected during the hardware test, they will be displayed at the
bottom left of the screen.
Each error must be confirmed by pressing the MFK. Then the system switches
to the home screen (see page 115). If the error message "System locked"
appears, a serious error has occurred. The device should not be put into oper-
ation. In this case, contact your service provider. All error messages listed in
the Table 13 on page 180.
All three LEDs will glow white during the booting process. This allows the user
to determine if the LEDs are working correctly.