Chapter 4: Commissioning
CONNECTING A SINGLE LINE PATIENT CIRCUIT
If the CARAT II pro ventilator is operated with a single line patient circuit, it
must not be used as a life-supporting device.
Connect the tube circuit to the device according to Figure 14 or Figure 15.
PAT
IEN
T
C
G
F
A
B
H
E
D
A
Patient side connection
B
Expiration valve
C
Ventilation tube
D
Bacterial filter
E
Adapter for bacterial filter
F
Pressure measuring tube
G
Control tube
H
Air outlet
Figure 14:
Connecting a single line patient circuit with pressure measuring tube
Should you be using a tube circuit without pressure measuring tube (Figure 15),
the circuit in use must be calibrated (see page 55).
D
PAT
IEN
T
A
C
B
E
A
Bacterial filter
B
Expiration valve
C
Ventilation tube
E
Control tube
E
Air outlet
Figure 15:
Connecting a single line patient circuit without pressure measuring tube