PW3390 Power Analyzer Communication Command Instruction Manual Second Edition
69/107
Query Measurement Data
Syntax
Query
:MEASure? <Item 1>, <Item 2>, ....... ,<Item 63>,<Item 64>
Response
<Item 1>, <Item 2>, ....... ,<Item 63>,<Item 64>
<Item 1 to 64>
: Fundamental measurement item arbitrary within
the parameter.
(
Refer to
4. Fundamental Measurement Item Parameters
Description
Query
Item Specified Mode (with parameters)
<Item> Describing any part will change it to this mode.
<Item> Create measurement data specified by this. A maximum of 64 items
is possible.
<Item> Can be arbitrarily specified within the fundamental measurement item
parameter, but anything else that is specified will return a command error.
<Item> Order is arbitrary. Creates the data in the order specified.
Item No Specified Mode (with no parameters)
<Item> If not part is specified, the mode will change to this.
Create measurement data of item specified under a type
of :MEASure:ITEM Command.
The measurement data order is fixed. (Refer to the output item and order
table)
Status data is always attached to the beginning of the data.
Example
Query
:MEAS? Urms1, P1, DEG1
(
Example of Item specified mode
)
Query the voltage RMS, effective power value, and power phase angle of
Channel 1.
Response
Urms1 151.63E+00,P1 5.74E+00,DEG1 83.80E+00 (WHEN HEADER
IS ON)
151.78E+00,5.58E+00,84.00E+00 (WHEN HEADER IS OFF)
Measurement data format
General
Measurement value
6 digits including digits after the decimal point
Index 2 digits
Integration Value
7 digits including digits after the decimal point
Index 2 digits
Time
Year,Month,Date
Hour,Minute,Second
Elapsed time (Hour,Minute,Second)
Elapsed Time (ms)
Error
Input over +9999.9E + 99
Note
Controls whether to omit the + at the beginning of the fraction and the preceding 0, under
the :TRANsmit:COLumnCommand.
If nothing is specified, the beginning of the fraction and the preceding 0 will be omitted
(:TRANsmit:COLumn 0)