PW3390 Power Analyzer Communication Command Instruction Manual Second Edition
75/107
Select and Query Harmonic List Communication Output Items
Syntax
Command
:MEASure:ITEM:HARMonic:LIST <LevelUI>,<LevelP>,
<Content UI>,<Content P>,<Phase Angle UI>,<Phase Angle P>
Query
:MEASure:ITEM:HARMonic:LIST?
Response
<Level UI>,<Level P>,<Content UI>,<Content P>,
<Phase Angle UI>,<Phase Angle P>
128
64
32
16
8
4
2
1
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
Level UI
HI4
HI3
HI2
HI1
HU4
HU3
HU2
HU1
Level P
FREQ
HP123
HP34
HP12
HP4
HP3
HP2
HP1
Content UI
HI4
HI3
HI2
HI1
HU4
HU3
HU2
HU1
Content P
-
HP123
HP34
HP12
HP4
HP3
HP2
HP1
Phase Angle UI
HI4
HI3
HI2
HI1
HU4
HU3
HU2
HU1
Phase Angle P
-
HP123
HP34
HP12
HP4
HP3
HP2
HP1
Description
Command
Sets the default send item of the response data in the default
mode as a response to ":MEASure:HARMonic?" as 0
– 255.
Specify the harmonic list (level, content, phase angle) here.
To set the item, execute the ON/OFF bit as shown above and specify with 0
–
255.
Query
Returns the setting for the default send item of the response data in the
default mode as a response to ":MEASure:HARMonic?" as 0
– 255.
Example
Command
:MEAS:ITEM:HARM:LIST 1,1,1,1,1,1
Set the U1,P1 level, Content, Phase Angle as the default output.
Query
:MEAS:ITEM:HARM:LIST?
Response
:MEASURE:ITEM:HARMONIC:LIST 1,1,1,1,1,1 (WHEN HEADER IS ON)
1,1,1,1,1,1 (WHEN HEADER IS OFF)
Note
Set the saving order with the ":MEASure:ITEM:HARMonic:ORDer"Command.