Select and Query Frequency Measurement Source
Syntax
Command
:FREQuency:SOURce[CH] <U/I>
Query
:FREQuency:SOIURce[CH]?
[CH]
・・・・
1,2,3,4
Response
U
: frequency source is voltage
I
: frequency source is current
Description
Command
Sets the specified channel’s frequency measurement source.
Query
Returns
the specified channel’s frequency measurement source
setting as a character string.
Example
Command
:FREQ:SOUR1 U
Select voltage as the frequency measurement source for Channel 1.
Query
:FREQ:SOUR1?
Response
:FREQUENCY:SOURCE1 U (WHEN HEADER IS ON)
U (WHEN HEADER IS OFF)
Select and Query Harmonic Synchronized Source
Syntax
Command
:HARMonic:SOURce <Synchronized source>
Query
:HARMonic:SOURce?
Response
<synchronized source>
:U1/U2/U3/U4/I1/I2/I3/I4/DC50ms/DC100ms/Ext
Description
Command
Sets the harmonic synchronized source.
Query
Returns the harmonic synchronized source setting as a character
string.
Example
Command
:HARM:SOUR U1
Set the harmonic synchronized source setting to U1.
Query
:HARM:SOUR?
Response
:HARMONIC:SOURCE U1 (WHEN HEADER IS ON)
U1 (WHEN HEADER IS OFF)
Note
Only the key of Ext can be set when the motor analysis function is implemented and CHB is set
as pulse.