PW3390 Power Analyzer Communication Command Instruction Manual Second Edition
27/107
Set and Query Phase Correction Angle for Current Sensors
Syntax
Command
:CURRent[CH]:DEGRee <Phase correction angle>
Query
:CURRent[CH]:DEGRee?
[CH]
・・・・
1,2,3,4
Response
Phase correction angle (°)
-90.00 to +90.00
Description
Command
Sets phase correction angle for current sensors. [CH]: 1 to 4.
Query
Returns setting for phase correction angle for current sensors in a string.
Example
Command
:CURR1:DEGR 90.00
Query
:CURR1:DEGR?
Response
:CURRENT1:90.00 (when HEADER is ON)
+90.00 (when HEADER is OFF)
Set and Query Phase Correction Frequency for Current Sensors
Syntax
Command
:CURRent[CH]:FREQuency <Correction frequency (kHz)>
Query
:CURRent[CH]:FREQuency?
[CH]
・・・・
1,2,3,4
Response
Correction frequency (kHz)
000.001 to 999.999
Description
Command
Sets phase correction frequency for current sensors. [CH]: 1 to 4.
Query
Returns setting for phase correction frequency for current sensors in a string.
Example
Command
:CURR1:FREQ 200.000
Query
:CURR1:FREQ?
Response
:CURRENT1:FREQUENCY 200.000 (when HEADER is ON)
200.000 (when HEADER is OFF)