PW3390 Power Analyzer Communication Command Instruction Manual Second Edition
58/107
Select and Query Noise Analysis Sampling Speed
Syntax
Command
:FFT:SAMPling <Sampling>
Query
:FFT : SAMPling?
Response
<Sampling>
:
<10kS/25kS/50kS/100kS/250kS/500kS>
Description
Command
Sets the noise analysis sampling speed.
Query
Returns the noise analysis sampling speed as a character string.
Example
Command
:FFT:SAMP 10kS
Set the noise analysis sampling speed to 10kS/s.
Query
:FFT:SAMP?
Response
:FFT:SAMPLING 10kS (WHEN HEADER IS ON)
10kS (WHEN HEADER IS OFF)
Note
Depending on the noise analysis sampling speed to be set, the noise lower limit frequency
may change automatically.
Set and Query Noise Analysis Window Function
Syntax
Command
:FFT:WINDow <0/1/2>
Query
:FFT:WINDow?
Response
0
:Rectangular
1
:Hanning
2
:Flat top
Description
Command
Sets the noise analysis window function.
Query
Returns the noise analysis window function setting as a numerical value.
Example
Command
:FFT:WIND 0
Set the noise analysis window function to rectangular.
Query
:FFT:WIND?
Response
:FFT:WINDOW 0 (WHEN HEADER IS ON)
0 (WHEN HEADER IS OFF)
Summary of Contents for PW3390