PW3390 Power Analyzer Communication Command Instruction Manual Second Edition
16/107
Set and Query D/A Output Items
Syntax
Command
:AOU
T:ITEM “Item 1”, “Item 2”,.......,”Item 16”
Query
:AOUT:ITEM?
Response
“Item 1”, “Item 2”, “Item 3”,.......,”Item 15”, “Item 16”
“Item 1 – 16” = Basic measured item parameters (Refer to
Description
Command
Sets the D/A output item. Output items can be specified from 1 to 16.
Output items in the unspecified D/A channels will not be changed.
Query
Sets the D/A output items and return them as words.
Example
Command
:AOUT:ITEM Urms1,Irms1,P1,Q1,S1,PF1
Set the D/A outputs from Channel 1 in sequence as Voltage CH1 RMs,
Current CH1 RMS, Effective Current CH1, Ineffective Current CH1, Apparent
Power CH1, and Power Factor Ch1.
Query
:AOUT:ITEM?
Response
:AOUT:ITEM
Urms1,Irms1,P1,Q1,S1,PF1,OFF,OFF,OFF,OFF,OFF,OFF
,OFF,OFF,OFF,OFF (when HEADER is ON)
Urms1,Irms1,P1,Q1,S1,PF1,OFF,OFF,OFF,OFF,OFF,OFF,OFF,OFF,OFF
,OFF (when HEADER is OFF) (when HEADER is OFF)
Select and Query Waveform Output
Syntax
Command
:AOUT:MONitor <ON/OFF>
Query
:AOUT:MONitor?
Response
ON
: Waveform output ON
OFF
: Waveform output OFF
Description
Command
Sets the Waveform output ON/OFF.
Query
Returns the setting of the waveform output as ON or OFF.
Example
Command
:AOUT:MON ON
Set the Monitor output as ON.
Query
:AOUT:MON?
Response
:AOUT:MONITOR ON (when HEADER is ON)
ON (when HEADER is OFF)
Summary of Contents for PW3390