PW3390 Power Analyzer Communication Command Instruction Manual Second Edition
63/107
Set and Query Integration Mode
Syntax
Command
:INTEGrate:MODE[CH] <DC/RMS>
Query
:INTEGrate:MODE[CH]?
[CH]
・・・・
1,2,3,4
Response
DC
: Integration DC mode
RMS
: Integration RMS mode
Description
Command
Sets the integration mode.
Query
Returns the integration mode setting as DC or RMS.
Example
Command
:INTEG:MODE1 DC
Set the integration mode of Channel 1 to DC.
Query
:INTEG:MODE1?
Response
:INTEGRATE:MODE1 DC (WHEN HEADER IS ON)
DC (WHEN HEADER IS OFF)
Note
By combining measurement lines (for above IP3W), the settings for integration modes of other
channels which are combined are also changed.
To get the DC mode, the current sensor used for the IP2W wiring must be the AC/DC type.
Execute Integration Data Reset
Syntax
Command
:INTEGrate:RESet
Description
Command
Resets integration data. Same operation as DATA RESET key of
the main instrument.
Example
Command
:INTEG:RES
Reset the integration data.
Note
This command is only available when the integration is in STOP status.
After sending the STOP command, transmit it with an interval of 250 ms
(1 sec when auto save ON).