PW3390 Power Analyzer Communication Command Instruction Manual Second Edition
29/107
Set and Query Current Range
Syntax
Command
:CURRent[CH]:RANGe <Current Range(NR2)>
Query
:CURRent[CH]:RANGe?
[CH]
・・・・
1,2,3,4
Response
0.1/0.2/0.4/0.5/0.8/1/2/4/5/8/10/20/40/50/80/
100/200/400/500/800/1000/2000/4000/8000/20000
Description
Command
Specifies the current range suitable for the sensor. (Unit is [A])
Query
Queries the current range setting. Returns the current range as a numerical
value in NR2 format.
Example
Command
:CURR1:RANG 1.0
Set the current Channel 1 to 1A range.
Query
:CURR1:RANG?
Response
:CURRENT1:RANGE 1.0 (WHEN HEADER IS ON)
1.0 (WHEN HEADER IS OFF)
Note
• Do not add a unit to the measurement range.
• Wait until the internal circuit has stabilized after changing a range before reading
measurements.
• When the range is specified, the auto range of the specified channel will be OFF.
• By combining measurement lines (for above IP3W), the settings for the current ranges of
other channels which are combined are also changed.