PW3390 Power Analyzer Communication Command Instruction Manual Second Edition
86/107
Setting and Querying Formula for Three-phase Power
Syntax
Command
:OPERation <TYPE1/TYPE2>
Query
:OPERation?
Response
TYPE1: 3P3W3M standard calculation method
TYPE2: Calculation method to improve the mutual compatibility with 3V3A
wiring of other power meter
Description
Command
Sets formulas of three-phase power.
Query
Returns formula settings of three-phase power as character string.
Example
Command
:OPER TYPE1
Set the formulas of three-phase power to TYPE1.
Query
:OPER?
Response
:OPERATION TYPE1 (WHEN HEADER IS ON)
TYPE1 (WHEN HEADER IS OFF)
Set and Query Execution Confirmation Message
Syntax
Command
:RS232c:ANSWer <ON/OFF>
Query
:RS232c:ANSWer?
Response
ON: Execution Confirmation Message
OFF: No Execution Confirmation Message
Description
Command
Sets the existence of execution confirmation message.
When there is a message, response will return even when the command is
being sent, and an execution confirmation message will be attached at the
end of the query’s response during query. The execution confirmation
message has a 3-
digit “nnn” format and shows an error at the nnnth
Command.
“000” is shown when there is an error.
Query
Returns the setting for the execution confirmation message as ON or OFF.
Example
Command
:RS232:ANSW ON
Set the execution confirmation message to ON.
Query
:RS232:ANSW?
Response
:RS232C:ANSWER ON;000 (WHEN HEADER IS ON)
ON,000 (WHEN HEADER IS OFF)
Note
The command name is RS232C, but is the same operation as when communicating with LAN
or USB.